Investigating the use of an octupole ion guide for ion storage and high-pass mass filtering to improve the quantitative performance of electrospray ion trap mass spectrometry

✍ Scribed by Robert D. Voyksner; Heewon Lee


Publisher
John Wiley and Sons
Year
1999
Tongue
English
Weight
148 KB
Volume
13
Category
Article
ISSN
0951-4198

No coin nor oath required. For personal study only.

✦ Synopsis


An octupole radio frequency (rf) ion guide was evaluated for storage and filtering of ions generated by electrospray prior to introduction into an ion trap mass spectrometer. The control of the rf potential on the ion guide enabled its operation as a high-pass mass filter, removing low-mass chemical noise that would normally fill the ion trap and result in reduced sensitivity, mass resolution, and dynamic range. Also, the ion guide can serve to store the high-pass filtered ions during the ion trap mass analysis, enabling nearly 100% duty cycle and reduction of the cycle time by a factor of two. The linearity, precision, and detection limits of the liquid chromatography (LC) electrospray -ion guide-ion trap MS/MS system were evaluated for the determination of ceftiofur in milk. A linear calibration (linear correlation coefficient of 0.986) from 2-200 ppb was obtained with a relative standard deviation for replicate analysis of less than 8%. A quantitation detection limit of 100 pg of ceftiofur on-column (2 ppb) was achieved from a milk extract. This detection limit is nearly a factor of 10 lower compared with the determination on the same electrospray ion trap system not using a rf ion guide for high-pass mass filtering and ion storage.


πŸ“œ SIMILAR VOLUMES


Positive ion electrospray ionization tan
✍ Robert L. St. Claire III πŸ“‚ Article πŸ“… 2000 πŸ› John Wiley and Sons 🌐 English βš– 107 KB πŸ‘ 1 views

A novel analytical procedure has been developed for the analysis of intracellular nucleotide triphosphates. Positive ion electrospray ionization tandem mass spectrometry (MS/MS) was interfaced to ion-pairing highperformance liquid chromatography (HPLC) utilizing a mobile phase containing 10 mM ammon

The use of high-flow high performance li
✍ M. Jemal; Yuan-Qing; Daisy B. Whigan πŸ“‚ Article πŸ“… 1998 πŸ› John Wiley and Sons 🌐 English βš– 210 KB πŸ‘ 1 views

Two bioanalytical methods have been developed and validated utilizing high flow high performance liquid chromatography (HPLC) for on-line purification of plasma and serum samples and electrospray tandem mass spectrometry for detection and quantitation. Each plasma or serum sample, after mixing with

High performance liquid chromatography m
✍ Mohammed Jemal; Zheng Ouyang; Deborah S. Teitz πŸ“‚ Article πŸ“… 1998 πŸ› John Wiley and Sons 🌐 English βš– 100 KB πŸ‘ 2 views

A systematic investigation was undertaken to study the effects of varying concentrations of additives in the acetonitrile/water high performance liquid chromatography mobile phase, especially formic acid and ammonium formate, on the negative ion electrospray response of a carboxylic acid compound. T

The Use of On-line Capillary Electrophor
✍ Michael X. Li; Jing-Tao Wu; Lin Liu; David M. Lubman πŸ“‚ Article πŸ“… 1997 πŸ› John Wiley and Sons 🌐 English βš– 204 KB πŸ‘ 2 views

Capillary electrophoresis/electrospray ionization using an ion trap storage/reflectron time-of-flight mass spectrometer detector (CE/ESI-IT/reTOF) is used to provide a rapid and sensitive method for analyzing structural variants in the hemoglobin (Hb) Ξ²-chain. The Hb Ξ±-and Ξ²-chains are separated and