𝔖 Bobbio Scriptorium
✦   LIBER   ✦

A small high-irradiance laser ionization time-of-flight mass spectrometer

✍ Scribed by Jian He; Rongfu Huang; Quan Yu; Yiming Lin; Wei Hang; Benli Huang


Publisher
John Wiley and Sons
Year
2009
Tongue
English
Weight
163 KB
Volume
44
Category
Article
ISSN
1076-5174

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

A small high‐irradiance laser ionization time‐of‐flight mass spectrometer (LI‐TOFMS) with orthogonal sample introduction was described. High irradiance of 6 × 10^10^ W/cm^2^ at 532 nm from a Nd : YAG laser was applied in the experiment to get a high ionization degree in plasma and to dissociate the interferential polyatomic ions. Meanwhile, the interferential multiply charged ions resulted by high‐irradiance were nearly eliminated in the spectrum by utilizing helium as the buffer gas in the ion source due to three‐body recombination, which resulted in a relatively clean background. Improved signal stability was obtained by automated step moving of the sample stage in short time intervals. By using two sets of Einzel lens in transport system, nearly uniform relative sensitivity coefficients (RSCs) were achieved for most of metal elements including light ions which were detected in extremely low sensitivity in previous hexapole transportation instrument. The resolving power reaches 2200, and the detection limits (DLs) are 10^−6^ g/g for metal elements in the steel standard. Copyright © 2009 John Wiley & Sons, Ltd.


📜 SIMILAR VOLUMES


A high resolving power multiple reflecti
✍ C. K. G. Piyadasa; P. Håkansson; T. R. Ariyaratne 📂 Article 📅 1999 🏛 John Wiley and Sons 🌐 English ⚖ 93 KB 👁 2 views

Two electrostatic mirrors, mounted symmetrically on the same optical axis facing each other, are used to increase the time-of-flight of molecular ions produced in matrix-assisted laser desorption/ionization (MALDI). The mirrors, which are used in the non-compensating mode, are located between a MALD

A miniaturized matrix-assisted laser des
✍ Robert D. English; Robert J. Cotter 📂 Article 📅 2003 🏛 John Wiley and Sons 🌐 English ⚖ 238 KB

## Abstract Mass correlated acceleration (MCA) has now been integrated into a 4 in (10.2 cm) matrix‐assisted laser desorption/ionization time‐of‐flight (MALDI‐TOF) mass spectrometer to achieve high resolving power across a broader mass range, without sacrificing detection of higher mass ions. The g

Application of a multi-turn time-of-flig
✍ Daisuke Okumura; Michisato Toyoda; Morio Ishihara; Itsuo Katakuse 📂 Article 📅 2004 🏛 John Wiley and Sons 🌐 English ⚖ 300 KB

## Abstract The circuit shape of the ion path, or the multi‐turn, provides a solution for achieving unrestricted mass resolution from time‐of‐flight mass analyzers. The potential of a multi‐turn type mass spectrometer, the MULTUM II, with a 1.308 m circuit controlled by four toroidal electric secto

Development of a tandem time-of-flight m
✍ Hirofumi Nagao; Shuichi Shimma; Shigeo Hayakawa; Kunio Awazu; Michisato Toyoda 📂 Article 📅 2010 🏛 John Wiley and Sons 🌐 English ⚖ 357 KB

## Abstract A new tandem time‐of‐flight mass spectrometer with an electrospray ionization ion source ‘ESI‐TOF/quadTOF’ was designed and constructed to achieve the desired aim of structural elucidation via high‐energy collision‐induced dissociation (CID), and the simultaneous detection of all fragme