## Abstract Tensile strength of the interface between hydroxyapatite (HA) and bone was tested. Scanning electron microscopy was used to observe the tensile failure mode and the morphological change of hydroxyapatite ceramic surface in bone. The porosity of hydroxyapatite is 14% and pore size less t
Mechanical and morphologic investigation of the tensile strength of a bone-hydroxyapatite interface
โ Scribed by Edwards, J. T. ;Brunski, J. B. ;Higuchi, H. W.
- Publisher
- John Wiley and Sons
- Year
- 1997
- Tongue
- English
- Weight
- 875 KB
- Volume
- 36
- Category
- Article
- ISSN
- 0021-9304
No coin nor oath required. For personal study only.
โฆ Synopsis
For load-bearing calcium-phosphate biomaterials, it is important to understand the relative contributions of direct physical-chemical bonding vs. mechanical interlocking to interfacial strength. In the limit of a perfectly smooth hydroxyapatite (HA) surface, a tensile test of the bone-HA interface affords an opportunity to isolate the bonding contribution related to HA surface chemistry alone. This study measured the bone-HA interfacial tensile strength for highly polished (approximately 0.05 micron alumina) dense HA disks (5.25 mm in diameter, 1.3 in mm thickness) in rabbit tibiae. Each of five rabbits received four HA disks, two per proximal tibia. Pull-off loads ranged from 3.14 +/- 2.38N at 55 days after implantation to 18.35 +/- 11.9N at 88 days; nominal interfacial tensile strengths were 0.15 +/- 0.11 MPa and 0.85 +/- 0.55 MPa, respectively. SEM of failed interfaces revealed failures between HA and bone, within the HA itself and within adjacent bone. Tissue remnants on HA were identified as mineralized bone with either a lamellar or trabecular structure. Oriented collagen fibers in the bone intricately interdigitated with the HA surface, which frequently showed breakdown at material grain boundaries and a rougher surface than originally implanted. Mechanical interlocking could not be eliminated as a mode of tissue attachment and contribution to bone-HA bonding, even after implanting an extremely smooth HA surface.
๐ SIMILAR VOLUMES
The interface shear strength of uncoated Ti-6A1-4V, dense sintered hydroxyapatite (HA), and HA-coated Ti-6A1-4V were compared. Interface shear strength was determined using a transcortical push-out model in dogs 4 and 12 weeks after implantation. The interface shear strength of dense sintered HA and
We performed a transcortical push-out test to determine the effect of surface roughness of hydroxyapatite (HA)coated implants on bone-implant shear strength in a canine model. Hydroxyapatite-and alumina-coated SUS316L with the same surface roughness (roughness average: Ra = 5 pm) and HA-coated Ti-6A
We evaluated the microstructural characteristics of newly formed bone tissue at the interface with cement. The bonecement interfaces of the femoral components of nine hip prostheses retrieved after loosening were investigated by means of X-ray diffraction on microareas and microhardness. The bone fa
A method of testing the tensile strength of the porcelain-gold bond has been developed. Tensile strengths of the bond are improved primarily by the formation of an oxide layer on the gold before baking the porcelain and not by the degree of gross surface roughness. The oxide layer may act either as
The purpose of this investigation was to determine the relationship between the degree of degeneration at the supraspinatus insertion, the tensile strength, and the site of failure of this tendon. Thirty-three fresh cadaveric shoulders (average age: 62 years; range: 39-83 years) were examined. A ten