We evaluated an immunoturbidimetric quantitation for serum myoglobin by the latex agglutination method using an automated biochemical analyzer. This method is rapid, specific, accurate, precise, and has wide dynamic range. The total assay time is 10 min and is performed at 37ยฐC with continuous monitoring at 570 nm.

The assay results were compared with radioisotopic immunoassay results and showed a good correlation coefficient, r = 0.99; Y = 0.98 x + 9.3; N = 79. Sera from healthy adults has a myoglobin concentration in the range of 15-80 ng/ml(N = 362). Sex-and age-related differences were observed. The serum myoglobin levels in males and elderly people showed higher concentration than in females and younger people. The peak elevation of serum myoglobin compared with other cardiac markers was observed within 6 hours after onset of chest pain as well as the CK-isoform ratio (MM3/MM,). All of the serum from 21 patients with definite acute myocardial infarction showed increased serum myoglobin levels (100-1200 ng/ml) upon admission and within 6 hours.

The results suggest that assays for serum myoglobin levels are helpful in the early diagnosis of acute myocardial necrosis.


๐Ÿ“œ SIMILAR VOLUMES


Measurement of serum ฮฒ2-microglobulin by
โœ F. Iguaz; J. Naval; L. Borque ๐Ÿ“‚ Article ๐Ÿ“… 1992 ๐Ÿ› Elsevier Science ๐ŸŒ English โš– 370 KB

A rapid and simple technique for the measurement of beta 2microglobulin on the Behring nephelometer was developed using a commercially available latex-anti [32-microglobulin. This is a fully automated assay and no pretreatment of sample is necessary. An excellent correlation with radioimmunoassay (r

Measurement of serum Lp(a) by COBAS MIRA
โœ Seiichi Fujita; Takahiro Sano; Yoshiaki Katayama ๐Ÿ“‚ Article ๐Ÿ“… 1994 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 308 KB

We had an opportunity to test a reagent kit for serum lipoprotein (a) [Lp(a)] assay, which was based on the latex immunoturbidimetric assay method, and applied it to COBAS MIRA (Roche). Reproducibility of the assay procedure was 0 . 9 2 4 0 % CV of three samples which contained 12.58-60.4 mgldl of L

Assessment of lectin activity in a toxic
โœ Aregheore, Eroarome M; Makkar, Harinder P S; Becker, Klaus ๐Ÿ“‚ Article ๐Ÿ“… 1998 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 178 KB ๐Ÿ‘ 1 views

Lectin activity in a toxic (Cape Verde) and non-toxic (Mexico) variety of Jatropha curcas seed meal was investigated using latex agglutination and haemagglutination assays. Lectin activity expressed as reciprocal of the minimum quantity (in mg) of Jatropha meal per ml of the assay mixture which prod