Spironolactone and potassium canrenoate (aldosterone antagonist diuretics) are often used with digoxin in clinical practice. Spironolactone, potassium canrenoate, and their common metabolite canrenone cross-react with the fluorescence polarization immunoassay (FPIA) for digoxin, and can falsely elev
Measurement of serum Lp(a) by COBAS MIRA using a latex immunoturbidimetric assay kit
โ Scribed by Seiichi Fujita; Takahiro Sano; Yoshiaki Katayama
- Publisher
- John Wiley and Sons
- Year
- 1994
- Tongue
- English
- Weight
- 308 KB
- Volume
- 8
- Category
- Article
- ISSN
- 0887-8013
No coin nor oath required. For personal study only.
โฆ Synopsis
We had an opportunity to test a reagent kit for serum lipoprotein (a) [Lp(a)] assay, which was based on the latex immunoturbidimetric assay method, and applied it to COBAS MIRA (Roche). Reproducibility of the assay procedure was 0 . 9 2 4 0 % CV of three samples which contained 12.58-60.4 mgldl of Lp(a). Minimum detectable concentration was 1.5-2.0 mg/dl. It was confirmed that in-terference of co-existing substances, i.e. hemoglobin, bilirubin, and intralipid, was negligible. No cross reactivity was seen with plasminogen. Correlation with a ELISA method was excellent. Frequency distribution of Lp(a) in healthy Japanese was similar to that found in white population.