Multielement analysis by means of plasma source mass spectrometry has evolved over the past decade primarily through the use of the quadrupole mass analyzer. In the present research a time-of-flight mass spectrometer (TOFMS) has been designed, built, and tested with an inductively coupled plasma (IC
Isotope ratios and abundance sensitivity obtained with an inductively coupled plasma-time-of-flight mass spectrometer
β Scribed by D.P. Myers; P.P. Mahoney; G. Li; G.M. Hieftje
- Publisher
- Elsevier Science
- Year
- 1995
- Tongue
- English
- Weight
- 877 KB
- Volume
- 6
- Category
- Article
- ISSN
- 1044-0305
No coin nor oath required. For personal study only.
β¦ Synopsis
Isotope ratios and abundance sensitivities have been determined with an inductively coupled plasma-time-of-flight mass spectrometer (ICP-TOFMS). Abundance sensitivities are at least in the 10(6) range for low abundance ions that precede high abundance ions. Three methods of detection for isotope-ratio measurement have been compared. The three systems involve gated detection followed by analog integration, analog averaging, or ion counting. Gated ion counting offers excellent precision-between 0. 64 and 1. 00% relative standard deviation (RSD). These values approach those predicted from counting statistics and are comparable to those reported for other inductively coupled plasma-mass spectrometry (ICP-MS) instruments. In addition, a greater number of accumulated counts or longer analysis times would afford precisions of 0. 1% with stable gating electronics. The accuracy of the counting method is in the 1-10% range if no correction for mass bias is performed. However, this ion counting method suffers from a limited dynamic range due to pulse pileup. Constant-fraction discrimination gated integration and commercial boxcar averager techniques offer a broader dynamic range because of their analog nature, but the attainable RSD values are limited by drift in the detection systems and by the methods employed to calculate an accurate ratio. Overall, mass bias in the ICP-TOFMS is more severe than previous work in ICP-MS due primarily to detection system bias.
π SIMILAR VOLUMES
An inductively coupled plasma-time-of-flight mass spectrometer (ICP-TOFMS) has been constructed and evaluated for elemental analysis. The instrument produces analog spectra similar to those from quadrupole inductively coupled plasma mass spectrometers. The large abundance of Ar ions is deflected awa
## Abstract CE was coupled to inductively coupled plasma MS (ICPβMS) and ESIβMS to identify and quantify the arsenic species arsenobetaine (AsB), arsenite (As^III^), arsenate (As^V^), and dimethylarsinic acid (DMA). A GCβflame ionization detector (FID)βbased German standard method and ICPβMS were u
Total arsenic concentrations and the concentrations of individual arsenic compounds were determined in liver samples of pinnipeds [nine ringed seals (Phoca hispida), one bearded seal (Erginathus barbatus)] and cetaceans [two pilot whales (Globicephalus melas), one beluga whale (Deliphinapterus leucu
We describe the coupling of a microfabricated fluidic device to an electrospray ionization (ESI) quadrupole time-of-flight mass spectrometer (QqTOFMS) for the identification of protein samples. The microfabricated devices consisted of three reservoirs connected via channels to a main capillary, whic