Intramedullary implant of plasma-sprayed hydroxyapatite coating: An interface study
β Scribed by Yang, C. Y. ;Wang, B. C. ;Lee, T. M. ;Chang, E. ;Chang, G. L.
- Publisher
- John Wiley and Sons
- Year
- 1997
- Tongue
- English
- Weight
- 536 KB
- Volume
- 36
- Category
- Article
- ISSN
- 0021-9304
No coin nor oath required. For personal study only.
β¦ Synopsis
An intramedullary implant model in the canine femora was developed to evaluate the mechanical and histological responses between cancellous bone and plasma-sprayed hydroxyapatite coatings (HACs) on Ti-6A1-4V implants, with 12-and 24-week follow-ups. HACs of different thicknesses were investigated. Results of the mechanical testings revealed that after 24 weeks of implantation, the mean shear strength (2.49 Β± 0.12 MPa) of the 50 m HACs was significantly higher (p < 0.05) than that of the 200 m HACs (1.44 Β± 0.19 MPa). However, using backscattered electron images (BEIs) throughout all the implant periods, no substantial histological variations in the extent of new bone apposition between the two HACs were observed. Occasionally, solution-mediated disintegration of the 50 m HAC was found 24 weeks postimplantation. Histomorphometric studies from the BEIs demonstrated that for both HACs the percentage of the direct HAC-cancellous bone contact was approximately 50% at 12 weeks and 75% at 24 weeks. After the mechanical tests, the 200 m HACs had fracture sites either inside the coating layers or at the HAC-titanium interfaces, which might explain why the mechanical performance of the 200 m HACs was inferior to that of the 50 m HACs even though both HACs had the same histological behaviors.
π SIMILAR VOLUMES
Tensile strength testing of the interface between bone and a plasma-sprayed hydroxyapatite (HA) coating-Ti-6Al-4V (implant I) and an HA coating-Ti (implant II) was performed. The bone-implant interface and tensile failure mode were evaluated by light microscopy (LM), SEM-energy dispersive X-ray anal
## Abstract We have established the electrical polarization method of hydroxyapatite (HA) coatings (HAC) for clinical use, such as dental and orthopedic implants. The HAC examined in the current study was prepared in titanium substrates by plasma spraying of Ξ²βtricalcium phosphate (TCP) powders fol
## Abstract The clinical use of plasmaβsprayed hydroxyapatite (HA) coatings on metal implants has been widely adopted because the HA coating can achieve the firmly and directly biological fixation with the surrounding bone tissue. However, the longβterm mechanical properties of HA coatings has been
This study was undertaken to evaluate the effect of coating characteristics on the mechanical strengths of the plasma-sprayed HA-coated Ti-6Al-4V implant system both in vitro and in vivo. Two types of HA coatings (HACs) with quite different microstructures, concentrations of impurityphases, and indi
Study of an amorphous phase in plasma-sprayed hydroxyapatite (HA) coatings is important owing to its unique characteristics and nonnegligible amount of the amorphous phase compared to crystalline HA. However, little is known about the component parts of an amorphous phase. It is known that amorphous