Comparison of plasma-sprayed hydroxyapatite coatings and zirconia-reinforced hydroxyapatite composite coatings:In vivo study
✍ Scribed by Lee, T. M. ;Yang, C. Y. ;Chang, E. ;Tsai, R. S.
- Publisher
- John Wiley and Sons
- Year
- 2004
- Tongue
- English
- Weight
- 268 KB
- Volume
- 71A
- Category
- Article
- ISSN
- 0021-9304
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
The clinical use of plasma‐sprayed hydroxyapatite (HA) coatings on metal implants has been widely adopted because the HA coating can achieve the firmly and directly biological fixation with the surrounding bone tissue. However, the long‐term mechanical properties of HA coatings has been concern for the long‐term clinical application. Previous research showed that the concept of adding ZrO~2~ as second phase to HA significantly increased the bonding strength of plasma‐sprayed composite material. The present work aimed to explore the biological properties, including the histological responses and shear strength, between the plasma‐sprayed HA and HA/ZrO~2~ coating, using the transcortical implant model in the femora of canines. After 6 and 12 weeks of implantation, the HA coating revealed the direct bone‐to‐coating contact by the backscattered electron images (BEIs) of scanning electron microscope (SEM), but the osseointegration was not observed at the surface of HA/ZrO~2~ coating. For new bone healing index (NBHI) and apposition index (AI), the values for HA implants were significantly higher than that for HA/ZrO~2~ coatings throughout all implant periods. After push‐out test, the shear strength of HA‐coated implants were statistically higher than HA/ZrO~2~ coated implants at 6‐ and 12‐week implantation, and the failure mode of HA/ZrO~2~ coating was observed at the coating‐bone interface by SEM. The results indicate that the firm fixation between bone and HA/ZrO~2~ has not been achieved even after 12‐week implantation. Consequently, the addition of ZrO~2~ could improve the mechanical properties of coatings, while the biocompatibility was influenced by the different material characteristics of HA/ZrO~2~ coating compared to HA coatings. © 2004 Wiley Periodicals, Inc. J Biomed Mater Res 71A: 652–660, 2004
📜 SIMILAR VOLUMES
## Abstract This study was undertaken to evaluate the effect of post‐hydrothermal treatment on the biological responses of the plasma‐sprayed hydroxyapatite (HA)‐coated Ti‐6Al‐4V implant system both __in vitro__ and __in vivo__. After hydrothermal treatment, the HA coating (HAC) shows the high mech
This study was undertaken to evaluate the effect of coating characteristics on the mechanical strengths of the plasma-sprayed HA-coated Ti-6Al-4V implant system both in vitro and in vivo. Two types of HA coatings (HACs) with quite different microstructures, concentrations of impurityphases, and indi
Study of an amorphous phase in plasma-sprayed hydroxyapatite (HA) coatings is important owing to its unique characteristics and nonnegligible amount of the amorphous phase compared to crystalline HA. However, little is known about the component parts of an amorphous phase. It is known that amorphous
## Abstract Bioceramic coatings like hydroxyapatite (HA) have shown promising bioactive properties in load‐bearing implant applications. The aim of this work is to deposit functionally graded HA/Ti layers consisting of an underlying Ti bond coat, the alternating layer, and an HA top‐layer on Ti6Al4
## Abstract One of the serious postoperative complications associated with joint replacement is bacterial infection. In addressing this problem, we have previously described the development of a novel thermal spraying technology combining silver (Ag) showing antibacterial activity with hydroxyapati