Focused transducer beams in tissue-mimicking material
โ Scribed by James A. Zagzebski; Richard A. Banjavic; Ernest L. Madsen; Mark Schwabe
- Publisher
- John Wiley and Sons
- Year
- 1982
- Tongue
- English
- Weight
- 740 KB
- Volume
- 10
- Category
- Article
- ISSN
- 0091-2751
No coin nor oath required. For personal study only.
โฆ Synopsis
Abstract
A comparison is made between pulse echo response profiles obtained in water and profiles obtained in a tissueโmimicking (TM) material for clinical ultrasound transducers. The TM material is assumed to provide a more accurate indication of actual beam profiles in soft tissue because of the presence of frequencyโdependent attenuation in the material. Pulse echo response profiles measured in the TM medium demonstrate a reduction in the focal distance and degraded resolution distal to the focal distance, in comparison to profiles obtained in water.