The analytical performance of a platinum preconcentration method using a 0.01 M HNO, carrier solution and an alumina microcolumn is discussed. On-line preconcentration is followed by flame atomic absorption spectrometry (FAAS), and off-line preconcentration by graphite furnace atomic absorption spec
Determination of Ultratrace Amounts of Copper and Cadmium in Seawater by Graphite Furnace Atomic Absorption Spectrometry with Flow-Injection Semi On-line Preconcentration
β Scribed by M.H. Wang; A.I. Yuzefovsky; R.G. Michel
- Publisher
- Elsevier Science
- Year
- 1993
- Tongue
- English
- Weight
- 827 KB
- Volume
- 48
- Category
- Article
- ISSN
- 0026-265X
No coin nor oath required. For personal study only.
β¦ Synopsis
Methods are described for the determination of ultratrace amounts of copper and cadmium in seawater by graphite furnace atomic absorption spectrometry with flow-injection, microcolumn preconcentration. A new type of (\mathrm{C}{18}) column loaded with sodium diethyldithiocarbamate (sodium-DDC) was used to extract copper and cadmium from seawater as the DDC chelates. The analytical effects of the (\mathrm{pH}) of the mixture of the sample and sodiumDDC solutions and the concentration of the chelating reagent were studied. Sodium-DDCloaded columns and unloaded (\mathrm{C}{18}) columns with different shapes and volumes were compared. To determine copper in seawater, a simple aqueous calibration was made with a mixture of palladium and magnesium nitrate as a matrix modifier, while for cadmium no matrix modifier was necessary. This method required only small seawater volumes, 600 and (400 \mu) for the determination of copper and cadmium respectively, with preconcentration factors of 15 -fold for copper and 10 -fold for cadmium. Detection limits for the preconcentration of aqueous solutions of copper and cadmium were 0.024 and (0.004 \mu \mathrm{g} \operatorname{liter}^{-1}(3 \sigma)), respectively. Results for determinations of copper and cadmium in National Research Council of Canada, CASS-2, Nearshore Seawater Reference Material showed no significant differences between the certified values and the measured values, based on Student's (t) test at the (95 %) confidence level. The relative standard deviations of the various measurements varied between 2 and 8%. O 1993 Academic Press, Inc.
π SIMILAR VOLUMES
Resins synthesized from qumohn-b-01 and resorcmol or hydroqmnone, with furfuraldehyde, formaldehyde or benzaldehyde as cross-hnkmg agent, were used for the preconcentrafion of nanogram amounts of lead The rate of exchange and activation energy of lead exchange were calculated Column separations of l
1,10 Phenanthroline (phen) has been tested as a complexing agent, which is more stable and selective than sodium-diethyldithiocarbamate (DDTC) and ammonium-pyrrolidinedithiocarbamate (APDC), for on-line preconcentration of copper on RP-C 18 material in a microcolumn with a flow injection-flame atomi
A flow injection sorbent extraction system was developed for determination of trace and ultratrace cadmium, copper and lead in biological and environmental standard reference materials by graphite furnace atomic absorption spectrometty (GFAAS). Using ammonium diethyldithiophosphate (DDPA) as complex
A flow injection on-line coprecipitation preconcentration system with diethyldithiocarbamate (DDTC) chelate of copper used as the coprecipitate carrier was coupled with flame atomic absorption spectrometry (FAAS) for the determination of trace silver. Silver was on-line coprecipitated with DDTC-Cu(I
Flow injection was used to develop an on-line automated focused microwave assisted procedure for the mineralization of adipose tissue and the determination of iron and zinc by graphite furnace atomic absorption spectrometry. Mineralization of samples was accomplished with a mixture of sulfuric and n