## Abstract To compare liver lesion detection rates, tissue signal and noise data, and qualitative parameters for breathhold (BH) and non‐breath‐hold (NBH) hybrid rapid acquisition with relaxation enhancement (RARE) and conventional spin‐echo (CSE) T2‐weighted (CSE‐T2) MR sequences, 20 patients wer
Comparison of breath-hold T1-weighted MR sequences for imaging of the liver
✍ Scribed by Richard C. Semelka; Ann Bagley Willms; Mark A. Brown; Elizabeth D. Brown; J. Paul Finn
- Publisher
- John Wiley and Sons
- Year
- 1994
- Tongue
- English
- Weight
- 910 KB
- Volume
- 4
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Three rapid T1‐weighted gradient‐echo techniques for imaging of the liver were compared: fast low‐angle shot (FLASH) and section‐selective (SSTF) and non‐section‐selective (NSTF) inversion‐recovery TurboFLASH. Ten healthy volunteers were imaged at 1.5 T, with breath‐hold images acquired in the transaxial and coronal planes and non‐breath‐hold images in the transaxial plane. Breath‐hold images were evaluated quantitatively and qualitatively, and non–breath‐hold images were evaluated qualitatively. FLASH images had significantly higher (P <.001) spleenliver signal difference–to‐noise ratios (SD/Ns) than NSTF and SSTF images. Liver signal‐to‐noise ratios (S/Ns) were significantly higher (P <.001) on FLASH images than on NSTF and SSTF images. With breath hold. FLASH images were rated as having the highest quality in the axial plane, followed by NSTF and SSTF images. In the coronal plane, NSTF images were rated as having the highest quality. For images acquired during patient respiration, NSTF images had the highest quality and showed the least degradation. The results suggest that FLASH images have the highest SD/N and S/N for liver imaging and have the highest quality in the axial plane. In patients who cannot suspend respiration, NSTF images may be least affected by breathing artifact and provide reasonable image quality.
📜 SIMILAR VOLUMES
## Abstract The potential of new high‐field‐strength magnetic resonance (MR) imaging sequences to evaluate bowel disease was investigated and compared with computed tomographic (CT) studies. Thirty‐two patients were studied, 14 with known or suspected gastrointestinal tumors and 18 with known or su
## Abstract ## Purpose To evaluate image quality and accuracy for the detection of focal hepatic lesions depicted on T2‐weighted images obtained with two high‐resolution free‐breathing techniques (navigator‐triggered turbo spin‐echo [TSE] and respiratory‐triggered TSE) and two standard‐resolution
## Abstract Twenty consecutive cancer patients with a solitary hepatic metastasis detected with dynamic contrast‐material—enhanced computed tomography (CT) who were considered for hepatic resection underwent magnetic resonance (MR) imaging within 18 days after CT. Histologic confirmation was obtain
## Abstract The purpose of this study was to evaluate the usefulness of multishot echo‐planar imaging in detecting liver tumors in comparison with respiratory triggered T2‐weighted fast‐spin‐echo (FSE) imaging. Thirty‐two patients with 70 focal liver lesions were imaged using a 1.5‐T high speed MR