✦ Synopsis


Abstract

Background

Bilateral control enables an operator to be placed away from the working environment. A slave manipulator usually works at a remote site or in a hazardous environment while the master operator senses the environment through the slave manipulator. Bilateral control can be used in surgeries. Usually doctors are very sensitive to the tactile sensation they feel through medical tools. If bilateral control is used in a surgery, ideally the doctor should perceive the sensation of both the body tissue and the reaction force from the specialized tool. Many surgical instruments are essentially similar at the tool tip while showing some structural differences at the gripping end. Therefore, it is advantageous for an expensive tool to be replaced by a virtual tool together with a simple tool. With the virtual tool, the surgeon should be able to perceive the feeling of the actual tool together with the reaction force coming from the body tissue; this may also reduce the time taken for a tool change during the surgery.

Methods

In this study, above ideas were implemented in a bilaterally controlled forceps robot and experiments were carried out to test the applicability of the concept. Bilateral control is used as the method of tactile feedback. Disturbance observer is used for robust control. Reaction force observer is used in place of a force sensor. Virtual model is used in place of the special tool properties.

Results

A virtual spring acting at the end of the forceps, was implemented to a bilateral control system, so that the operator at the master side could perceive the feeling of the spring together with reaction forces from the surgical environment. Operator could feel the vivid sensation of the environment through the master manipulator.

Conclusions

The virtual model adds a load to the system. The master and slave responses are slightly different, but the result is good enough for the application. This novel approach could be used in many master–slave applications. Copyright Β© 2007 John Wiley & Sons, Ltd.


πŸ“œ SIMILAR VOLUMES


Optimizing port placement for robot-assi
✍ A. L. Trejos; R. V. Patel; I. Ross; B. Kiaii πŸ“‚ Article πŸ“… 2007 πŸ› Wiley (Robotic Publications) 🌐 English βš– 728 KB

## Abstract ## Background Proper placement of ports during robot‐assisted endoscopic surgery is critical to the success of the procedure. In current practice, port placement methods do not consider the ability of the robot to manoeuvre the tools. ## Methods This paper proposes to choose the best

A technical challenge for robot-assisted
✍ J Stallkamp; R D Schraft πŸ“‚ Article πŸ“… 2005 πŸ› Wiley (Robotic Publications) 🌐 English βš– 288 KB

## Abstract In minimally invasive surgery, a higher degree of accuracy is required by surgeons both for current and for future applications. This could be achieved using either a manipulator or a robot which would undertake selected tasks during surgery. However, a manually‐controlled manipulator c

EndoCAS navigator platform: a common pla
✍ Giuseppe Megali; Vincenzo Ferrari; Cinzia Freschi; Bruno Morabito; Filippo Caval πŸ“‚ Article πŸ“… 2008 πŸ› Wiley (Robotic Publications) 🌐 English βš– 571 KB

## Abstract ## Background Computer‐assisted surgery (CAS) systems are currently used in only a few surgical specialties: ear, nose and throat (ENT), neurosurgery and orthopaedics. Almost all of these systems have been developed as dedicated platforms and work on rigid anatomical structures. The de