Tissue response to implanted polymers: The significance of sample shape

✍ Scribed by Matlaga, Barbara F. ;Yasenchak, Lewis P. ;Salthouse, Thomas N.


Publisher
John Wiley and Sons
Year
1976
Tongue
English
Weight
366 KB
Volume
10
Category
Article
ISSN
0021-9304

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Studies were designed to demonstrate the need for standardization of shape of samples used as implants to evaluate histotoxicity of polymer materials. Six medical‐grade polymers (polypropylene, polyethylene, polyurethane, silicone rubber, poly(vinyl chloride), and Teflon) were extruded as rods with circular‐, triangular‐, and pentagonal‐shaped crosss ections, and were implanted in rat gluteal muscles for 14 days. Evaluation of the tissue response was assessed by quantitating cellular lysosomal acid phosphatase enzyme activity by using microspectrophotometry. All triangular‐shaped implants showed the highest enzyme activity and cellular response; pentagon shapes showed less, and circular rods showed the lowest activity. The results demonstrate the need for standard sample shape for valid comparative studies of tissue response to implanted polymers.


📜 SIMILAR VOLUMES


Biophysical evaluation of the tumorigeni
✍ Habal, Mutaz B. ;Powell, Rodger D. 📂 Article 📅 1980 🏛 John Wiley and Sons 🌐 English ⚖ 454 KB

## Abstract An evaluation of the biophysical changes of methylcholanthrene induced tumor (M~4~) in laboratory animals receiving four different alloplastic implants was conducted. The implants are silicone, polyether urethane, poly(methyl methacrylate), and bioglass. The tumor was allowed to grow to

An In Vivo Study of the Host Response to
✍ Alexandra P. Marques; Rui L. Reis; John A. Hunt 📂 Article 📅 2005 🏛 John Wiley and Sons 🌐 English ⚖ 473 KB 👁 1 views

## Abstract **Summary:** Implant failure is one of the major concerns in the biomaterials field. Several factors have been related to the fail but in general these biomaterials do not exhibit comparable physical, chemical or biological properties to natural tissues and ultimately, these devices can

In vivo time course studies of the tissu
✍ Patrick W. Stroman; Jean-Claude Dorvil; Yves Marois; Nicolas Poddevin; Robert Gu 📂 Article 📅 1999 🏛 John Wiley and Sons 🌐 English ⚖ 164 KB 👁 1 views

Magnetic resonance (MR) imaging and relaxation time measurements of bioresorbable implants made of polylactic acid (PLA), as well as the surrounding tissues, were carried out over a period of 6 months to monitor the implant state and the body's responses, and to determine how these processes are ref