In time-resolved contrast-enhanced 3D MR angiography, spatial resolution is traded for high temporal resolution. A hybrid method is presented that attempts to reduce this tradeoff in two of the spatial dimensions. It combines an undersampled projection acquisition in two dimensions with variable rat
Time-resolved contrast-enhanced carotid imaging using undersampled projection reconstruction acquisition
✍ Scribed by Jiang Du; Sean B. Fain; Frank R. Korosec; Thomas M. Grist; Charles A. Mistretta
- Publisher
- John Wiley and Sons
- Year
- 2007
- Tongue
- English
- Weight
- 691 KB
- Volume
- 25
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Purpose
To investigate the utility of nonuniform angular spacing of projections in a three‐dimensional (3D) hybrid undersampled projection reconstruction (PR) acquisition for contrast‐enhanced (CE) time‐resolved carotid imaging.
Materials and Methods
Carotid CE magnetic resonance angiography (CE‐MRA) was performed on seven healthy volunteers using a time‐resolved hybrid sequence that combined undersampled PR acquisition in‐plane and Cartesian slice encoding through‐plane. The undersampling streak artifact comes mainly from the superior–inferior (S/I) direction in carotid imaging and is suppressed by nonuniform distribution of the projections. Phantom and volunteer studies were performed to demonstrate its efficacy.
Results
The undersampling streak artifact was significantly suppressed through a nonuniform distribution of the projection angles with more projections aligned along the S/I direction. The hybrid PR sequence combined with nonuniform distribution of the projection angles provided time‐resolved images of the carotid arteries with high temporal resolution (two seconds per frame) and high spatial resolution (1.0 × 1.0 × 1.5 mm^3^) simultaneously.
Conclusion
High‐resolution dynamic imaging of the carotid arteries is feasible with the use of a hybrid undersampled PR acquisition. Undersampling streak artifact can be suppressed significantly through nonuniform distribution of the projections. J. Magn. Reson. Imaging 2007. © 2007 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract ## Purpose To investigate the application of time‐resolved vastly undersampled isotropic projection reconstruction (VIPR) in contrast‐enhanced magnetic resonance angiography of the distal extremity (single station), and peripheral run‐off vasculature in the abdomen, thigh, and calf (th
## Abstract The hybrid projection reconstruction (PR) imaging provides high temporal resolution through an undersampled PR acquisition for the in‐plane dimensions and Cartesian slice encoding for the through‐plane dimension. The undersampling of projection data introduces streak artifact, which may
Time-resolved contrast-enhanced 3D MR angiography (MRA) methods have gained in popularity but are still limited by the tradeoff between spatial and temporal resolution. A method is presented that greatly reduces this tradeoff by employing undersampled 3D projection reconstruction trajectories. The v
## Abstract ## Purpose To introduce a bilateral implementation of an adaptive imaging technique in which both dynamic and high resolution breast MR images are acquired simultaneously. ## Materials and Methods Adaptive three‐dimensional bilateral breast imaging in the sagittal plane was achieved
Imaging of the blood vessels below the knee using contrast-enhanced (CE) MRI is challenging due to the need to coordinate image acquisition and arrival of the contrast in the targeted vessels. Time-resolved acquisitions have been successful in consistently capturing images of the arterial phase of t