Thermoelectric enzyme sensor for measuring blood glucose
β Scribed by Michael J. Muehlbauer; Eric J. Guilbeau; Bruce C. Towe; Tedd A. Brandon
- Publisher
- Elsevier Science
- Year
- 1990
- Tongue
- English
- Weight
- 800 KB
- Volume
- 5
- Category
- Article
- ISSN
- 0956-5663
No coin nor oath required. For personal study only.
β¦ Synopsis
A new calorimetric sensor has been developed which employs a thin-film thermopile in association with an immobilized enzyme. The thermopile detects the minute temperature rise that occurs when a specific chemical substrate is catalyzed by the enzyme. A prototype sensor is described which generates an equivalent proportional voltage response to glucose concentrations present in either buffer solution or blood. These sensors have remained useful for up to 18 days when operated intermittently for measuring glucose in buffer solutions, or for up to 4 days when operated continuously. When implanted inside cardiovascular shunts on anesthetized dogs, the sensors responded appropriately to changes in the blood glucose concentration.
π SIMILAR VOLUMES
A glucose sensor intended for future use in diabetic patients undergoing major surgery is described. It involves glucose oxidase immobilised at a platinised activated carbon electrode (PACE). The sensor gave a nonoxygen dependent amperometric response in whole undiluted blood and did not require the
An enzyme electrode is described for glucose determination in unstirred, undiluted whole blood. The system comprises an H,O,-detecting electrode upon which is placed a membrane laminate incorporating glucose oxidase. The external membrane was pretreated with methyltrichlorosilane. The electrode res