A sequential injection analysis system for the determination of calcium in water, urine and pharmaceutical samples is proposed. The method is based on the fast complexation reaction between cresolphthalein complexone and calcium with spectrophotometric detection. Baseline shift due to a background o
Spectrophotometric determination of calcium in paper machine white water by sequential injection analysis
β Scribed by Johan Nyman; Ari Ivaska
- Publisher
- Elsevier Science
- Year
- 1995
- Tongue
- English
- Weight
- 639 KB
- Volume
- 308
- Category
- Article
- ISSN
- 0003-2670
No coin nor oath required. For personal study only.
β¦ Synopsis
A sequential injection analysis method is developed for the determination of calcium in white water of a paper machine.
The determination is based bn the formation of a coloured complex between Ca*+ and o-cresolphthalein complexone, which is monitored spectrophotometrically.
Parameters of the method were optimized in order to give a working range as broad as possible. A working range of S-500 mg I-' Ca*+ was achieved with a sample volume of 30 ~1. Preliminary tests of the method were conducted using off-line samples from different paper mills in Finland. The method developed was also tested in an on-line application at a paper mill which produces wood-containing paper grades.
π SIMILAR VOLUMES
A sequential injection analysis system for the determination of sulphate in natural waters and industrial effluents has been developed. The method is based on the turbidimetric determination of sulphate using barium chloride as reagent and measuring the absorbance of the formed suspension spectropho
A flow-mjectlon spectrophotometrx procedure 1s described for the sequential deternunatlon of glucose, fructose and sucrose m a sample solution by usmg s&e and dual lmmobdtzed enzyme reactors m a smgle mamfold Concentrations III the range 10-5-10-3 M are determmed urlth analysis rates of 30-40 determ
Flow-injection analysis has been applied for the determination of traces of calcium and magnesium in rare earth oxide samples. The method is based on the fluorescence of the calcein complexes of calcium and magnesium in a solution with high pH. The analyte ions are separated from the rare earth ion