## Abstract Examination of a great number of tissue samples taken from the newly formed capsules surrounding artificial joints reveals small particles of prosthetic material. Abraded from the joint by wear and tear, these particles of plastic, metal, and acrylic cement initiate a foreignβbody react
Simulation of polyethylene wear in ankle joint prostheses
β Scribed by Carol J. Bell; John Fisher
- Publisher
- John Wiley and Sons
- Year
- 2007
- Tongue
- English
- Weight
- 293 KB
- Volume
- 81B
- Category
- Article
- ISSN
- 1552-4973
No coin nor oath required. For personal study only.
β¦ Synopsis
Abstract
The performance of total ankle replacements (TARs) have not been comparable to those of the other major joints of the lower extremity. The aim of this work was to develop a new simulator test to compare the wear of a new mobile bearing TAR (Mobility) with one with a good clinical history, the Buechel Pappas, using kinematic inputs derived from the literature. The wear rate for the Mobility components was lower than that for the BuechelβPappas ankle joints at all time points. The wear rate for both sets of components increased with the inclusion of an anterior/posterior displacement in the kinematic inputs. This was expected as the components are subjected to higher kinematic demands and reproduces similar effects found in knee prostheses. This study has demonstrated that it is possible to study wear of TARs in a modified simulator originally designed for total knee replacements. It was also shown that the new Mobility ankle compares favorably with the Buechel Pappas ankle, which has a successful clinical history, under the simulator test conditions described. Β© 2006 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2006
π SIMILAR VOLUMES
The aim of this article is to present a numerical model of wear. The original research is that this model is local in space and time and deals with dynamic and large strains applications. The model is based on the Archard's law and coupled with contact and friction conditions. A numerical treatment
Frictional heat was monitored during wear tests of ultrahigh molecular weight polyethylene acetabular cups bearing against femoral balls of metal or ceramic in a hip simulator, using bovine serum as a lubricant. About 1 to 2 h of continuous cycling were required for the temperature in the zone of co
Differences in bearing surface conformity and wear mechanisms suggest that the polyethylene (PE) wear debris generated by total knee replacement (TKR) prostheses should be different than that in total hip replacement prostheses (THR). To address this issue, PE wear debris and the cellular response i