We present the design and preliminary results from a Fourier transform ion cyclotron resonance (ICR) mass spectrometer developed for the direct detection of W/visible laser-induced fluorescence of trapped, massselected, gas-phase ions. A 3 T superconducting magnet and an open-ended multi-section cyl
Sharing a Pair of Electrodes between Excitation and Detection in a Fourier Transform Ion Cyclotron Resonance Mass Spectrometer
✍ Scribed by Gérard Mauclaire; Joël Lemaire; Michel Heninger
- Publisher
- John Wiley and Sons
- Year
- 1996
- Tongue
- English
- Weight
- 298 KB
- Volume
- 10
- Category
- Article
- ISSN
- 0951-4198
No coin nor oath required. For personal study only.
✦ Synopsis
Time-sharing of electrodes between excitation and detection for Fourier transform (lT) study of the coherent cyclotron motion of ions is discussed, together with the choice of a suitable commutation device. A practical implementation is described, and the application to an open-end, flat, ion cyclotron resonance cell presented. In addition, the switching technique described is useful to minimise ion losses due to resistive damping at the magnetron frequencies, and could also be used in conventional FT ion cyclotron resonance cells for quadrupolar excitation and detection, and for ion axialization.
📜 SIMILAR VOLUMES
An external source 7 T Fourier transform ion cyclotron resonance (FTICR) mass spectrometer offers three main novel features. First, a 9-way ion-source cross allows for mounting of up to three ionization sources simultaneously, thereby minimizing 'downtime' for changing ion sources. Second, an electr
Electron capture dissociation of the peptide Substance P is reported for the first time, with an unmodified, commercially available Fourier transform ion cyclotron resonance (FTICR) mass spectrometer. The fragmentation pattern is compared with that obtained with collisionally induced dissociation of
We present Fourier transform ion cyclotron resonance (FTICR) mass spectra at a magnetic field of 20 T; more than twice the highest field previously used for FTICR. Our instrument is based on a resistive magnet installed at the National High Magnetic Field Laboratory. The magnet has a 50mm diameter b