Real-time cardiac MRI at 3 tesla
✍ Scribed by Krishna S. Nayak; Charles H. Cunningham; Juan M. Santos; John M. Pauly
- Publisher
- John Wiley and Sons
- Year
- 2004
- Tongue
- English
- Weight
- 446 KB
- Volume
- 51
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Real‐time cardiac and coronary MRI at 1.5T is relatively “signal starved” and the 3T platform is attractive for its immediate factor of two increase in magnetization. Cardiac imaging at 3T, however, is both subtly and significantly different from imaging at 1.5T because of increased susceptibility artifacts, differences in tissue relaxation, and RF homogeneity issues. New RF excitation and pulse sequence designs are presented which deal with the fat‐suppression requirements and off‐resonance issues at 3T. Real‐time cardiac imaging at 3T is demonstrated with high blood SNR, blood‐myocardium CNR, resolution, and image quality, using new spectral‐spatial RF pulses and fast spiral gradient echo pulse sequences. The proposed sequence achieves 1.5 mm in‐plane resolution over a 20 cm FOV, with a 5.52 mm measured slice thickness and 32 dB of lipid suppression. Complete images are acquired every 120 ms and are reconstructed and displayed at 24 frames/sec using a sliding window. Results from healthy volunteers show improved image quality, a 53% improvement in blood SNR efficiency, and a 232% improvement in blood‐myocardium CNR efficiency compared to 1.5T. Magn Reson Med 51:655–660, 2004. © 2004 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract The sodium concentration gradient in the kidney (from the cortex to the medulla) serves to regulate fluid homeostasis and is tightly coupled to renal function. It was previously shown that renal function and pathophysiology can be characterized in rat kidneys by measuring the sodium gra
## Abstract A hybridized dual‐imaging system combining real‐time ultrasound imaging and MRI was utilized for cardiac imaging at 1.5 T and 3 T. The ultrasound scanner with a programmable software interface was connected via computer to the MRI scanner. Electronic noise was eliminated with electromag
## Abstract ## Purpose: To describe a rapid T2\*‐weighted (T2\*W), three‐dimensional (3D) echo planar imaging (EPI) sequence and its application in mapping local magnetic susceptibility variations in 3 Tesla (T) prostate MRI. To compare the sensitivity of T2\*W EPI with routinely used T1‐weighted