Quantification of T2 in the abdomen at 3.0 T using a T2-prepared balanced turbo field echo sequence
✍ Scribed by Caroline L. Hoad; Eleanor F. Cox; Penny A. Gowland
- Publisher
- John Wiley and Sons
- Year
- 2010
- Tongue
- English
- Weight
- 431 KB
- Volume
- 63
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
The T~2~‐prepared balanced turbo field echo sequence has been used to measure T~2~ in phantoms and in vivo in the abdomen with low sensitivity to radiofrequency pulse errors. The effects of noise, errors in the pulse flip angles, and off resonance effects on the results have been simulated. It was found from simulations that for normal conditions, including the flip angle in the fit improved the systematic errors due to radiofrequency pulse errors and noise in the results to less than 1% without significantly increasing the random errors. For a 0.3% noise level, the standard deviation in the measured T~2~ was approximately 0.003 ms. Off‐resonance effects had a minimal effect on the measured T~2~ value. The T~2~ at 3.0 T of various abdominal organs was measured, in particular the liver (31 ± 6 ms), spleen (54 ± 15 ms), kidney cortex (76 ± 6 ms), kidney medulla (61 ± 8 ms), and pancreas (42 ± 20 ms). Magn Reson Med, 2010. © 2009 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract ## Purpose To evaluate a free‐breathing navigator triggered T2‐weighted turbo spin‐echo sequence with prospective acquisition correction (T2w‐PACE‐TSE) for MRI of the upper abdomen in comparison to a conventional T2‐weighted TSE (T2w‐CTSE), a single‐shot TSE (T2w‐HASTE), and a T1‐weigh
## Abstract In vivo quantitation of hepatic iron content is useful in diagnosis and staging of several iron related diseases. We used an experimental model of hepatic iron overload to determine the correlation between iron content and T2 relaxation time in rat liver. Experiments were carried out at
## Abstract ## Purpose: To evaluate motion artifacts, tissue contrasts, and lesion detectability in the neck with the periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) (BLADE) technique. ## Materials and Methods: A total of 46 patients referred for MRI of t
Isothermal Ageing in the T Range from 0 "C to 180 "C of a Technical A1-2.0 at.% Zn-1.3 at.% Mg Alloy after Direct Quench The variations of the structure and the mechanical properties had been investigated by means of TEM investigations and determination of the microhardness number in the T, range b