On the evaluation of ion source conditions for optimum sensitivity of chemical ionization
β Scribed by R. P. Morgan; E. J. Hayward; G. Steel
- Publisher
- John Wiley and Sons
- Year
- 1979
- Tongue
- English
- Weight
- 207 KB
- Volume
- 14
- Category
- Article
- ISSN
- 1076-5174
No coin nor oath required. For personal study only.
β¦ Synopsis
Abstract
The best operating conditions that yield maximum sensitivity at minimum source pressure have been evaluated for the chemical ionization source in a Kratos MSβ25 mass spectrometer. The best reagent gas pressure is shown to be different for each gas and is significantly lower than is usually reported in the literature.