## Abstract Heparin and low molecular mass heparin (LMMH) consists of complex mixtures of sulphated linear oligosaccharides that are difficult to analyse. An on‐line size exclusion chromatographic/electrospray ionization (ESI) mass spectrometric method that allows the determination of more than 60
On-line size-exclusion chromatography/electrospray ionisation mass spectrometry of aquatic humic and fulvic acids
✍ Scribed by Linn Persson; Tomas Alsberg; Gyula Kiss; Göran Odham
- Publisher
- John Wiley and Sons
- Year
- 2000
- Tongue
- English
- Weight
- 129 KB
- Volume
- 14
- Category
- Article
- ISSN
- 0951-4198
No coin nor oath required. For personal study only.
✦ Synopsis
Isolated aquatic humic and fulvic acids were analysed with on-line size exclusion chromatography/electrospray ionisation mass spectrometry (SEC/ESI-MS). An eluent composition which enabled electrospray ionisation was identified. The SEC separation improved interpretability of mass spectra and may open up new possibilities for molecular weight determination of humic substances. A linear dose-response relationship over a factor of 20 was obtained and the limit of detection was 50ng/uL for humic and fulvic acids. Spectral changes due to different ionisation conditions (pH and cone voltage) were investigated. A natural water sample from a Swedish lake was analysed. Copyright 2000 John Wiley & Sons, Ltd.