## Abstract ## Purpose: To assess personal time‐weighted average (TWA) static magnetic field and time‐varying magnetic field exposure for several system testing tasks routinely conducted by engineers at an MRI manufacturing plant. ## Materials and Methods: Personal exposure was measured using a
Occupational exposure to static and time-varying gradient magnetic fields in MR units
✍ Scribed by Jill K. Bradley; Matilda Nyekiova; David L. Price; Lindsay D'jon Lopez; Therese Crawley
- Publisher
- John Wiley and Sons
- Year
- 2007
- Tongue
- English
- Weight
- 203 KB
- Volume
- 26
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Purpose
To assess exposure of MR staff to static and time‐varying gradient magnetic fields by direct measurement and questionnaire and to compare these to the action values and exposure limits stated in published guidance.
Materials and Methods
Instantaneous and time‐averaged exposure to static magnetic fields were measured on clinical and research MR scanners. A questionnaire was sent to 100 clinical MR centers to estimate the average static field exposure. Instantaneous measurements of the time‐varying magnetic field were conducted on a 1.5T conventional and a 0.6T open‐bore MR system.
Results
The 24‐hour time‐averaged static field exposure was measured to be 1.8 ± 1 mT and the questionnaire produced a 24‐hour time‐average of 2.3 mT. The maximum static field strength recorded during patient set up was 554 mT. Time‐varying gradient field exposure was found to exceed the action values up to 30 cm from the bore entrance or magnet rim.
Conclusion
The time‐varying gradient field exposure limit is the only limit to be exceeded. The time‐averaged static field exposure to staff was found to be 100 times lower than the recommended guidelines. J. Magn. Reson. Imaging 2007;26:1204–1209. © 2007 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract ## Purpose To determine if magnetic field exposure close to two clinical 1.5T magnetic resonance imaging (MRI) scanners during image acquisition and when moving in the spatially‐varying static magnetic field is compliant with European Union (EU) Directive 2004/40/EC (the Directive). #
The effects of exposure to clinical magnetic resonance imaging (MRI) on analgesia induced by the mu opiate agonist, fentanyl, was examined in mice. During the dark period, adult male mice were exposed for 23.2 min to the time-varying (0.6 T/sec) magnetic field (TVMF) component of the MRI procedure.
The purpose of this study was to evaluate the influence on fetal cell growth in vitro of rapidly changing magnetic gradient fields such as those produced by the gradient coils of a typical magnetic resonance (MR) imager. The static magnetic field and the radiofrequency pulses were disabled during al
It has been proposed that chronic exposure to extremely low frequency (ELF) magnetic fields (MF) in occupational environments could represent a risk factor for a number of disorders. Medical and technical workers in hospitals have been reported to be exposed to relatively strong ELF fields. The pres