We have developed a new method for the integration of ion-sensitive membranes and silicon sensor chips. The ion-sensitive membrane is deposited in pyramidal containments produced on silicon by an anisotropic etching technique. Here we demonstrate the application of this method to the development of nitrate-sensitive sensors. PVC membranes are cast into the membrane containment. The sensors show an improved mechanical stability and lifetime compared to ISFETs. Slopes, detection limits and selectivities of the containment sensors are similar to those of ion-selective electrodes with internal electrolyte solution using the same membrane composition. The results of the drift and hysteresis measurements are sutficiently good for most practical applications.
๐ SIMILAR VOLUMES
A fast and convenient method is reported for preparing ion-selective electrode membranes based on bisphenol A epoxyacrylate, 1,6-hexanedioldiacrylate and benzophenone photoinitiator. A calcium-selective electrode is used as example; responses are Nemstian in the range 1O-'-lO-5 M with the calcium sa
This article reviews recent advances in the ยฎeld of potentiometric solvent polymeric membrane electrodes. These sensors have found widespread applications in a variety of ยฎelds, especially in the area of clinical diagnostics. Emphasis is given on the discussion of the theoretical and practical limit
Single-piece, conducting polymer-based, potentiometric sensors with enhanced cationic sensitivity were obtained by doping a polypyrrole with metal-complexing, niultivalent anions. Sulphosalycilic acid, Tiron, Eriochrome Black T and Kalces have been used as doping ions. Predominant sensitivity for co