𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Metallized sheathless electrospray emitters for use in capillary electrophoresis orthogonal time-of-flight mass spectrometry

✍ Scribed by David R. Barnidge; Stefan Nilsson; Karin E. Markides; Håkan Rapp; Klas Hjort


Publisher
John Wiley and Sons
Year
1999
Tongue
English
Weight
174 KB
Volume
13
Category
Article
ISSN
0951-4198

No coin nor oath required. For personal study only.

✦ Synopsis


Chromium-gold coated fused silica emitters for electrospray ionization have been fabricated with lifetimes upwards of 100 hours. The coating on these emitters is extremely durable withstanding electrical discharges, abrasion and mechanical stress. Usable flow rates range from 30 nL/min to 1 mL/min using both standard electrospray solutions and aqueous buffers. Coated capillary electrophoresis columns incorporating the emitter provide excellent separation, low limits of detection (70-700 attomoles for peptides), and are easy to use. Performance of the emitters in continuous infusion experiments and capillary electrophoresis/ electrospray ionization mass spectrometry applications are presented.


📜 SIMILAR VOLUMES


Capillary electrophoresis/electrospray i
✍ Ping Cao; Mehdi Moini 📂 Article 📅 1998 🏛 John Wiley and Sons 🌐 English ⚖ 151 KB 👁 2 views

Capillary electrophoresis/electrospray ionization (CE/ESI) high mass accuracy time-of-flight mass spectrometry was used for the first time to characterize small proteins using peptide mapping. To identify small proteins, the intact proteins were first analyzed to obtain their average molecular weigh

On-column conductive coating for thermol
✍ Stefan Nilsson; Karin E. Markides 📂 Article 📅 2000 🏛 John Wiley and Sons 🌐 English ⚖ 68 KB 👁 1 views

A simple and very mild approach for the application of a conductive layer for sheathless electrospray has been developed. A modified 'fairy dust' method is employed in which 2 mm gold particles are applied by a thin layer of silicone on shaped tips. This novel approach comprises fabrications at room

Accurate mass determinations for the con
✍ A. C. Hogenboom; W. M. A. Niessen; D. Little; U. A. Th. Brinkman 📂 Article 📅 1999 🏛 John Wiley and Sons 🌐 English ⚖ 113 KB 👁 2 views

The identification of polar microcontaminants in surface water is an important issue in environmental analysis. Liquid chromatography/mass spectrometry (LC/MS) is frequently applied for this purpose. However, even in combination with tandem mass spectrometry (MS/MS), unambiguous identification of th

The Use of On-line Capillary Electrophor
✍ Michael X. Li; Jing-Tao Wu; Lin Liu; David M. Lubman 📂 Article 📅 1997 🏛 John Wiley and Sons 🌐 English ⚖ 204 KB 👁 2 views

Capillary electrophoresis/electrospray ionization using an ion trap storage/reflectron time-of-flight mass spectrometer detector (CE/ESI-IT/reTOF) is used to provide a rapid and sensitive method for analyzing structural variants in the hemoglobin (Hb) β-chain. The Hb α-and β-chains are separated and