## Abstract Three‐dimensional __T__~1~‐weighted magnetization‐prepared rapid gradient‐echo (MP‐RAGE) sequences with centric phase encoding (PE) in the inner loop provide structural brain images with a high spatial resolution and high tissue contrast. A disadvantage of this sequence type is the susc
Inherent insensitivity to RF inhomogeneity in FLASH imaging
✍ Scribed by Danli Wang; Keith Heberlein; Stephen LaConte; Xiaoping Hu
- Publisher
- John Wiley and Sons
- Year
- 2004
- Tongue
- English
- Weight
- 368 KB
- Volume
- 52
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Radiofrequency (RF) field inhomogeneity is an unavoidable problem in MRI, and it becomes severe at high magnetic fields due to the dependence of B~1~ on the sample. It leads to nonuniformities in image intensity and contrast, causing difficulties in quantitative interpretation and image segmentation. In this work, it is observed that with the fast low‐angle shot (FLASH) sequence, which is often used for anatomic imaging and morphometric studies, sensitivity to RF inhomogeneity can be substantially reduced when the same coil is used for both transmission and reception, and an appropriate nominal flip angle is employed. This observation can help us understand the signal behavior of FLASH in the presence of RF inhomogeneity, and provide a guide for selecting parameters in FLASH imaging. Magn Reson Med 52:927–931, 2004. © 2004 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract Amide proton transfer (APT) imaging results in a representative human volunteer. a: B~0~ map calculated from the low SNR z‐spectrum images by the B~0~ correction algorithm. b: MTR~asym~ map and c: MTR~SAFARI~ map calculated from the high SNR APT images without B~0~ correction. The uncor
## Abstract Although RARE and GRASE can produce single‐shot images of excellent quality, their utility has been restricted because preparation of the magnetization with interesting contrast before imaging can cause severe artifacts. These artifacts relate to the strong sensitivity of multiple spin
Fears have been voiced that excessive tissue heating could occur in the event that first, a surface coil is placed with its axis parallel to the transmitting rffield leading to a maximal coupling of the two coils and second, the decoupling circuit of the surface coil breaks down. To avoid an rf coup
## Abstract ## Purpose: To reliably compensate bias field effects in abdominal areas to accurately quantify visceral adipose tissue using standard T1‐weighted sequences on MR scanners with up to 3 Tesla (T) field strength. ## Materials and Methods: Compensation is achieved in two steps: The bias