Abstract

Inflow effects were studied for T~1~‐weighted, fast gradient‐echo, contrast‐enhanced MRI. This was done on the basis of realistic simulations (e.g., taking slice profiles into account) for unsteady flow. The area under the point spread function (PSF) was used to estimate the flow‐related enhancement. A simple analytical model that accurately describes the inflow effects was derived and validated. This model was used to correct the experimental perfusion calibration curves (signal intensity vs. relaxation rate) for inflow effects. Hepatic perfusion measurements, performed on patients, were analyzed in terms of a dual‐input, first‐order linear model. It was shown that inflow causes incorrect perfusion input functions. The resulting estimated perfusion parameters displayed a systematic error of typically 30–40%. By performing two extra time‐resolved flow measurements during the examination, one can correct the input functions. Magn Reson Med 51:710–717, 2004. © 2004 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES


Precision in measurements of perfusion a
✍ Lucy E. Kershaw; David L. Buckley 📂 Article 📅 2006 🏛 John Wiley and Sons 🌐 English ⚖ 286 KB

## Abstract Dynamic contrast‐enhanced MRI is used to estimate microvascular parameters by tracer kinetics analysis. The time for the contrast agent to travel from the artery to the tissue of interest (bolus arrival time (BAT)) is an important parameter that must be measured in such studies because

MRI of bone tumors: Fast STIR imaging as
✍ Osamu Tokuda; Noriko Hayashi; Naofumi Matsunaga 📂 Article 📅 2004 🏛 John Wiley and Sons 🌐 English ⚖ 329 KB

## Abstract ## Purpose To compare the usefulness of short inversion recovery (STIR) and T1‐weighted, contrast‐enhanced, fat‐suppressed (T1W‐CEFS) sequences for the evaluation of bone tumors. ## Materials and Methods Eighteen patients with 19 bone tumors who underwent both STIR and T1W‐CEFS imagi

Tumor-liver contrast and subjective tumo
✍ Koichi Ishiyama; Manabu Hashimoto; Jun-ichi Izumi; Jiro Watarai; Satoshi Shibata 📂 Article 📅 2008 🏛 John Wiley and Sons 🌐 English ⚖ 227 KB

## Abstract ## Purpose To compare respiratory‐triggered T2‐weighted fast spin‐echo (RTT2W‐FSE) and gradient T2\*‐weighted recalled‐echo (T2\*W‐GRE) images for visualization of malignant hepatic tumors using ferucarbotran‐enhanced magnetic resonance (MR) imaging. ## Materials and Methods Ferucarb

Quantitative measurement of blood-brain
✍ Saeid Taheri; Charles Gasparovic; Nadim Jon Shah; Gary A. Rosenberg 📂 Article 📅 2010 🏛 John Wiley and Sons 🌐 English ⚖ 301 KB

## Abstract Breakdown of the blood‐brain barrier (BBB), occurring in many neurological diseases, has been difficult to measure noninvasively in humans. Dynamic contrast‐enhanced magnetic resonance imaging measures BBB permeability. However, important technical challenges remain and normative data f