## Abstract ## Purpose To combine parallel imaging with 3D single‐shot acquisition (echo volumar imaging, EVI) in order to acquire high temporal resolution volumar functional MRI (fMRI) data. ## Materials and Methods An improved EVI sequence was associated with parallel acquisition and field of
Improved echo volumar imaging (EVI) for functional MRI
✍ Scribed by Wietske van der Zwaag; Susan Francis; Richard Bowtell
- Book ID
- 102952881
- Publisher
- John Wiley and Sons
- Year
- 2006
- Tongue
- English
- Weight
- 601 KB
- Volume
- 56
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Echo volumar imaging (EVI) is a 3D modification of echo‐planar imaging (EPI) that allows data from an entire volume to be acquired following a single RF excitation. EVI provides a high rate of volumar data acquisition, which is advantageous for functional MRI (fMRI). However, few studies to date have applied EVI to fMRI, since because of gradient hardware limitations EVI generally has to be used with long sampling times, resulting in high sensitivity to susceptibility‐induced distortions. In this study we modified the EVI sequence to improve its suitability for fMRI. The sampling time is reduced by the use of a high gradient‐switching frequency, a small number of echoes, and outer volume suppression (OVS); rewind gradients ameliorate Nyquist ghosting; and phase correction via a calibration scan reduces ghosting and distortion. It is shown that the modified EVI sequence allows fMRI data to be acquired with a temporal resolution of 167 ms. Magn Reson Med, 2006. © 2006 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract Improvements in the echo‐volumar imaging (EVI) technique and its application in whole‐body studies are described. Using an in‐house built 0.5 T echo‐planar imaging system, complete modulus three‐dimensional image data are acquired in 102 ms with real‐time display. Hardware limitations h
## Abstract A 3‐dimensional MRI method has been developed for functional mapping of the human brain, based on blood oxygenation level dependent (BOLD) contrast mechanisms. The method uses recently introduced principles of echo‐shifted FLASH to acquire a single 3D data set in 20 s. The technique was
## Abstract The partial Fourier gradient‐echo echo planar imaging (EPI) technique makes it possible to acquire high‐resolution functional MRI (fMRI) data at an optimal echo time. This technique is especially important for fMRI studies at high magnetic fields, where the optimal echo time is short an