A series of experiments was conducted to evaluate the partition efficiency of the cross-axis synchronous flow-through coil planet centrifuge. The preliminary studies indicated that short coils (5 m) mounted on 15-and 25-cm diameter holders in either a central or a lateral location produce a high par
Improved cross-axis synchronous flow-through coil planet centrifuge for performing counter-current chromatography : II. Studies on retention of stationary phase in short coils and preparative separations in multilayer coils
โ Scribed by Molina Bhatnagar; Hisao Oka; Yoichiro Ito
- Publisher
- Elsevier Science
- Year
- 1989
- Tongue
- English
- Weight
- 977 KB
- Volume
- 463
- Category
- Article
- ISSN
- 1873-3778
No coin nor oath required. For personal study only.
โฆ Synopsis
Performance of the apparatus was evaluated in terms of stationary phase retention, partition efficiency and sample loading capacity. Preliminary studies with short coils revealed high retention of the stationary phase under a proper combination of the head-tail elution and planetary motion. Preparative capability of the apparatus was successfully demonstrated on efficient multigram separations of 2,4-dinitrophenyl amino acids, indole auxins, and bacitracin in a pair of large multilayer coils with a total capacity of 1.5 1.