## Abstract The aim of this study was to determine the relationships between the surface properties and blood compatibility of in‐use diamond‐like carbon (DLC) coatings for cardiovascular components. Commercially available DLC films were characterized with respect to surface topography and wettabil
Fluorinated diamond-like carbon as antithrombogenic coating for blood-contacting devices
✍ Scribed by Terumitsu Hasebe; Atsushi Shimada; Tetsuya Suzuki; Yoshiaki Matsuoka; Toshiya Saito; Satoshi Yohena; Aki Kamijo; Nobuyuki Shiraga; Mutsumi Higuchi; Kanako Kimura; Hirokuni Yoshimura; Sachio Kuribayashi
- Book ID
- 102873419
- Publisher
- John Wiley and Sons
- Year
- 2006
- Tongue
- English
- Weight
- 815 KB
- Volume
- 76A
- Category
- Article
- ISSN
- 1549-3296
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Diamond‐like carbon (DLC) is being considered for widespread clinical use as a surface coating for cardiovascular devices. We synthesized fluorinated DLC (F‐DLC) coatings in order to create a more hydrophobic surface with improved antithrombogenicity and flexibility when compared with conventional DLC coatings by combining the inertness of DLC films with the advantage of fluorination. The purpose of this study was to evaluate the in vitro hemocompatibility and in vivo biocompatibility of the F‐DLC coating for medical devices. The in vitro whole blood model confirmed that platelet loss was lower in the F‐DLC group than in the noncoated group (SUS316L), which suggests the adhesion of a smaller number of platelets to F‐DLC‐coated materials. Furthermore, the biomarkers of mechanically induced platelet activation (beta‐thromboglobulin) and activated coagulation (thrombin‐antithrombin‐three complex) were markedly reduced in the F‐DLC‐coated group. In vivo rat implant model studies revealed no excessive local and systemic inflammatory responses in the F‐DLC group. The thickness of the fibrous tissue capsule surrounding the F‐DLC‐coated disk was almost equal to that of the noncoated SUS316L disk, which has the favorable biocompatibility for metallic implant materials. F‐DLC coating thus appears to be a promising candidate for use as a coating material in blood‐contacting devices. © 2005 Wiley Periodicals, Inc. J Biomed Mater Res, 2006
📜 SIMILAR VOLUMES
Diamond-hke carbon (DLC) coated mlcroporous polycarbonate membranes were used as outer covering membranes m glucose enzyme electrodes With optlmlsed DLC deposltlon and the use of 0 01 pm mxroporous polycarbonate, blocompatabdlty has been seen to unprove with only a loss of 6% response after 30 mm ex