Effects of pulsed laser systems on stapes footplate

✍ Scribed by Jovanovic, Sergije; Schonfeld, Uwe; Prapavat, Viravuth; Berghaus, Alexander; Fischer, Rudi; Scherer, Hans; Muller, Gerhard J.


Publisher
John Wiley and Sons
Year
1997
Tongue
English
Weight
336 KB
Volume
21
Category
Article
ISSN
0196-8092

No coin nor oath required. For personal study only.

✦ Synopsis


Background and objective:

The aim of the present study was to investigate the tissue ablation capacity of various pulsed lasers at the stapes footplate.

Study design/materials and methods:

Isolated human stapes and bovine compact-bone platelets were used to determine the effective laser parameters and appropriate application technique for achieving a perforation measuring 500-600 microns in diameter. of interest were also the shape and quality of the perforations, the reproducibility of the perforation effect, and the thermically altered marginal zones occurring at the footplate. three pulsed laser systems were used: excimer, holmium:yag (ho:yag), and erbium:ysgg (er:ysgg) lasers.

Results:

The tissue-ablating effect of pulsed laser systems permits a precise and controlled management of the stapes footplate through low and readily reproducible ablation rates. the extent of thermic side effects at the footplate is lower in comparison to the purely thermically acting cw and superpulse laser systems. the er:ysgg laser exhibits the highest ablation rate at the stapes and is thus the most effective laser for perforation of the stapes footplate. though somewhat less effective, the ho:yag laser also appears to be suitable for stapedotomy. on the other hand, we do not consider the applied excimer laser (308 nm) to be particularly appropriate at the stapes because of its low ablation rates.

Conclusion:

Thus, the erbium laser could represent an alternative to the argon, ktp 532, and co2 lasers, already clinically successful in stapes surgery. however, further studies are necessary to examine the transmission of thermic energy into the vestibule and the acoustic stress to the inner ear during laser stapedotomy, to be able to make a definitive statement about the safest and most effective laser system for stapes surgery.


πŸ“œ SIMILAR VOLUMES


Thermic effects in the β€œvestibule” durin
✍ Jovanovic, S.; SchΓΆnfeld, U.; Fischer, R.; DΓΆring, M.; Prapavat, V.; MΓΌller, G.; πŸ“‚ Article πŸ“… 1998 πŸ› John Wiley and Sons 🌐 English βš– 253 KB πŸ‘ 1 views

## Background and Objective : Apart from the ablation properties at the stapes footplate, the degree of thermic loading in the inner ear is important in determining the suitability of pulsed lasers for stapedotomy. The aim of the study is to compare the thermic effects in the vestibule with differ

Effect of pulse duration on microsecond-
✍ Krishna M. Bhatta; Norman S. Nishioka πŸ“‚ Article πŸ“… 1989 πŸ› John Wiley and Sons 🌐 English βš– 348 KB

The pulsed dye laser (wavelength 504 nm, pulse duration 1 ps) is widely used for fragmenting urinary and biliary calculi. In this study, the performance of this laser was compared with pulsed dye lasers producing pulse durations of 8 and 20 ps. Fragmentation thresholds and fragmentation rates were m

The effect of epidermal pigmentation on
✍ Dr. O. T. Tan; Russell Kerschmann; John A. Parrish πŸ“‚ Article πŸ“… 1984 πŸ› John Wiley and Sons 🌐 English βš– 625 KB

The effect of epidermal pigmentation on the threshold exposure dose for inducing purpura with a tunable dye laser at 577 nm, 1.5 1 s pulse duration, was studied in 2 I human volunteers with varied genetically determined amounts of melanin. More laser energy was required to produce purpura as constit

Histologic comparison of the pulsed dye
✍ O. T. Tan; T. J. Stafford; S. Murray; A. K. Kurban πŸ“‚ Article πŸ“… 1990 πŸ› John Wiley and Sons 🌐 English βš– 862 KB

Albino pig skin was exposed to the copper vapor (CVL) and flashlamp pulsed dye (PDL) lasers at 578 nm with a 3 mm diameter spotsize over a range of fluences until purpura and whitening were first established. The total irradiation time was the parameter that was varied in order for the CVL to reach