Effect of Solvent on Dynamic Range and Sensitivity in Pneumatically-assisted Electrospray (Ion Spray) Mass Spectrometry

✍ Scribed by Risto Kostiainen; Andries P. Bruins


Publisher
John Wiley and Sons
Year
1996
Tongue
English
Weight
767 KB
Volume
10
Category
Article
ISSN
0951-4198

No coin nor oath required. For personal study only.

✦ Synopsis


Mass-analyzed detector signal and spray current have been measured in pneumatically-assisted electrospray mass spectrometry. The sample was tetrabutylammonium bromide dissolved in water, methanol, acetonitrile, chloroform, dichloromethane or toluene. At low sample concentrations ( S 5 x M) the ion signal rose with increasing sample concentration. Above lo-' M the ion signal was fixed and independent of sample concentration.

Comparison of signals with spray currents for tetrabutylammonium bromide at 2 x M in difterent solvents revealed a strong correlation between ion signal and spray current. Apparently, the abundance of the tetrabutylammonium ion at m/z 242 is fully controlled by the amount of charge on droplets, while other solvent properties such as volatility, surface tension and polarity do not play a role at low tetrabutylammonium bromide concentrations. Thus, water is a poor solvent for electrospray because it does not allow efficient droplet charging, not because it is less volatile and more difficult to spray than organic solvents. The ion signal at 2 x M tetrabutylammonium bromide in different solvents is highest for dichloromethane. At high sample concentrations (> M) the dependence of the ion signal on spray current is lost. It appears impossible to convert a high charge on droplets into sample ions. Creation of droplets having a surface fully covered with sample is assumed to be the cause of ion signal saturation. Increasing the sample concentration will only increase the number of ions inside a droplet. The number of sample ions at the surface escaping into the gas phase is k e d and independent of sample concentration above M.


πŸ“œ SIMILAR VOLUMES


Aerosol Matrix-assisted Laser Desorption
✍ Russell, David H.; Beeson, Michelle D. πŸ“‚ Article πŸ“… 1996 πŸ› John Wiley and Sons 🌐 English βš– 837 KB

The effects of methanol, ethanol, propan-2-01 and acetonitrile on the aerosol matrix-assisted laser desorptionlionization (MALDI) ion signal of hradykinin, bovine insulin and myoglobin were examined. The effect of these solvents mixed with water was also studied. The aerosol MALDI ion signal for eac

Sensitivity improvement in the analysis
✍ R. A. M. van der Hoeven; A. J. P. Hofte; U. R. Tjaden; J. van der Greef; Nelson πŸ“‚ Article πŸ“… 1998 πŸ› John Wiley and Sons 🌐 English βš– 172 KB πŸ‘ 2 views

The on-line coupling of high-performance anion-exchange chromatography (HPAEC) with ion spray mass spectrometry, for the analysis of neutral and acidic oligosaccharides, is described. On-line desalting is performed by a cation-exchange membrane system which replaces sodium for hydronium ions from th

A further study on the combined use of i
✍ Yuhki Tsukamoto; Tomofumi Santa; Hiroo Yoshida; Hiroshi Miyano; Takeshi Fukushim πŸ“‚ Article πŸ“… 2006 πŸ› John Wiley and Sons 🌐 English βš– 313 KB πŸ‘ 2 views

The combined use of a so-called internal standard and the isotope-labeled derivatization reagent for the quantification of analytes for liquid chromatography-mass spectrometry (LC/MS) was further studied. The sample solution (containing the analytes and an internal standard) was derivatized with the

Effect of combined changes in delayed ex
✍ Daniele Vitalini; Placido Mineo; Emilio Scamporrino πŸ“‚ Article πŸ“… 1999 πŸ› John Wiley and Sons 🌐 English βš– 169 KB πŸ‘ 3 views

Data reported here show that, in the delayed extraction matrix-assisted laser desorption/ionization time-offlight (DE-MALDI-TOF) mass spectrometric analysis of synthetic polydisperse polymers, different experimental conditions of spectral recording are required to optimize the signal in all the m/z