Matrix selection and sample preparation technique, namely finding a suitable matrix-analyte combination, has been one of the most important concerns in applying positive and negative ion matrix-assisted laser desorption/ionization (MALDI) mass spectrometry to synthetic polymers. Most matrices used i
Effect of combined changes in delayed extraction time and potential gradient on the mass resolution and ion discrimination in the analysis of polydisperse polymers and polymer blends by delayed extraction matrix-assisted laser desorption/ionization time-of-flight mass spectrometry
β Scribed by Daniele Vitalini; Placido Mineo; Emilio Scamporrino
- Publisher
- John Wiley and Sons
- Year
- 1999
- Tongue
- English
- Weight
- 169 KB
- Volume
- 13
- Category
- Article
- ISSN
- 0951-4198
No coin nor oath required. For personal study only.
β¦ Synopsis
Data reported here show that, in the delayed extraction matrix-assisted laser desorption/ionization time-offlight (DE-MALDI-TOF) mass spectrometric analysis of synthetic polydisperse polymers, different experimental conditions of spectral recording are required to optimize the signal in all the m/z regions of the spectrum. The effect of combined changes in delay time and grid voltage % values on both mass resolution and mass accuracy of DE spectra of a polyethylene glycol sample (PEG mix , with a well-defined molar distribution of its components) is discussed. The necessity of a compromise between the values of these two parameters is shown. Furthermore, the occurrence of analyte discrimination, which can invalidate the composition analysis especially in the case of polymer blends, is demonstrated.
π SIMILAR VOLUMES
Pulsed-delayed extraction using a simple high voltage transistor switch together with various sample purification approaches were used to enhance the resolution in matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOFMS) of oligonucleotides up to 60 bases long. This
The isomeric sugar-branched cyclodextrin (CD) derivatives of 6-O-glucosyl-betaCD (G1-betaCD) and 6,6-di-O-glucosyl-alphaCD (G1, G1-alphaCD) were analyzed by matrix-assisted laser desorption/ionization time-of-flight (MALDI-TOF) mass spectrometry at the different acceleration voltages of 20 and 8 kV.