๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

Effect of pressure on solute retention in the ionization control mode in reversed-phase liquid chromatography

โœ Scribed by Nobuo Tanaka; Tadanori Yoshimura; Mikio Araki


Publisher
Elsevier Science
Year
1987
Tongue
English
Weight
654 KB
Volume
406
Category
Article
ISSN
1873-3778

No coin nor oath required. For personal study only.

โœฆ Synopsis


The pressure in a chromatographic system was found to have a significant effect on the retention of solutes in the ionization control mode in reversed-phase liquid chromatography. An increase in the pressure in a column by 200 kg/cm2 resulted in as much as a 12% variation in k', and caused peak reversal with the variation of the pressure. The variation of retention due to the change in pressure was attributed to the shift of the pH of the mobile phase and a change in the extent of ionization of a solute. The direction and the magnitude of the pressure effect on retention, which depend on the combination of a solute and the type of buffer solution, can be predicted on the basis of the volume changes for ionization of a solute and a buffer species. It was also shown that the effect of pressure on retention can be avoided by the choice of a suitable buffer component in the mobile phase.


๐Ÿ“œ SIMILAR VOLUMES


Influence of pressure on solute retentio
โœ Victoria L. McGuffin; Christine E. Evans ๐Ÿ“‚ Article ๐Ÿ“… 1991 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 612 KB

## Abstract Although the mobile phase is generally considered to be incompressible in liquid chromatography, the local pressure appears to have a significant effect upon solute retention. In this investigation, laser fluorescence detection was utilized to measure solute retention directly along a r

Separation of isotopic compounds by reve
โœ Tanaka, N. ;Yamaguchi, A. ;Hashizume, K. ;Araki, M. ;Wada, A. ;Kimata, K. ๐Ÿ“‚ Article ๐Ÿ“… 1986 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 440 KB

Isotope separation utilizing ionization control in reversed-phase liquid chromatography was studied. Several organic acids and bases labeled with oxygen-18 and nitrogen-15 were successfully separated from unlabeled compounds by using either 6-m microbore columns or recycle chromatography with short