โฆ Synopsis
The pressure in a chromatographic system was found to have a significant effect on the retention of solutes in the ionization control mode in reversed-phase liquid chromatography. An increase in the pressure in a column by 200 kg/cm2 resulted in as much as a 12% variation in k', and caused peak reversal with the variation of the pressure. The variation of retention due to the change in pressure was attributed to the shift of the pH of the mobile phase and a change in the extent of ionization of a solute. The direction and the magnitude of the pressure effect on retention, which depend on the combination of a solute and the type of buffer solution, can be predicted on the basis of the volume changes for ionization of a solute and a buffer species. It was also shown that the effect of pressure on retention can be avoided by the choice of a suitable buffer component in the mobile phase.
๐ SIMILAR VOLUMES
## Abstract Although the mobile phase is generally considered to be incompressible in liquid chromatography, the local pressure appears to have a significant effect upon solute retention. In this investigation, laser fluorescence detection was utilized to measure solute retention directly along a r
Isotope separation utilizing ionization control in reversed-phase liquid chromatography was studied. Several organic acids and bases labeled with oxygen-18 and nitrogen-15 were successfully separated from unlabeled compounds by using either 6-m microbore columns or recycle chromatography with short