## ABSTRACT A highly sensitive, specific and simple LC‐MS/MS method was developed for the simultaneous estimation of dexlansoprazole (DEX) with 50 μL of human plasma using omeprazole as an internal standard (IS). The API‐4000 LC‐MS/MS was operated under multiple reaction‐monitoring mode using elect
Development and validation of a highly sensitive LC-MS/MS method for simultaneous quantitation of ethionamide and ethionamide sulfoxide in human plasma: application to a human pharmacokinetic study
✍ Scribed by Abhijeet Y. Deshpande; Sandip Gurav; Ravindra Punde; Vishal Zambre; Rahul Kulkarni; Sarvesh Pandey; Ashish Mungantiwar; Ramesh Mullangi
- Publisher
- John Wiley and Sons
- Year
- 2011
- Tongue
- English
- Weight
- 882 KB
- Volume
- 25
- Category
- Article
- ISSN
- 0269-3879
- DOI
- 10.1002/bmc.1554
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
A highly sensitive and specific LC‐MS/MS method has been developed for simultaneous quantification of ethionamide and ethionamide sulfoxide in human plasma (300 µL) using prothionamide as an internal standard (IS). Solid‐phase extraction was used to extract ethionamide, ethionamide sulfoxide and IS from human plasma. The chromatographic separation of ethionamide, ethionamide sulfoxide and IS was achieved with a mobile phase consisting of 0.1% acetic acid : acetonitrile (20:80, v/v) at a flow rate of 0.50 mL/min on a Peerless Basic C~18~ column. The total run time was 3.5 min and the elution of ethionamide, ethionamide sulfoxide and IS occurred at 2.50, 2.18 and 2.68 min, respectively. A linear response function was established for the range of concentrations 25.7–6120 ng/mL (r > 0.998) for ethionamide and 50.5–3030 ng/mL (r > 0.998) for ethionamide sulfoxide. The intra‐ and inter‐day precision values for ethionamide and ethionamide sulfoxide met the acceptance as per FDA guidelines. Ethionamide and ethionamide sulfoxide were stable in battery of stability studies, viz. bench‐top, autosampler and freeze–thaw cycles. The developed assay was applied to a pharmacokinetic study in humans. Copyright © 2011 John Wiley & Sons, Ltd.
📜 SIMILAR VOLUMES
## Abstract A highly sensitive and specific LC‐MS/MS method has been developed and validated for the estimation of pramipexole (PPX) with 500 µL human plasma using memantine as an internal standard (IS). The API‐4000 was operated under multiple‐reaction monitoring mode (MRM) using the electrospray