Determination of hydrogen peroxide by using a flow injection system with immobilized peroxidase and long pathlength capillary spectrophotometry
✍ Scribed by Alexandros Ch Pappas; Constantine D Stalikas; Yannis Ch Fiamegos; Miltiades I Karayannis
- Publisher
- Elsevier Science
- Year
- 2002
- Tongue
- English
- Weight
- 116 KB
- Volume
- 455
- Category
- Article
- ISSN
- 0003-2670
No coin nor oath required. For personal study only.
✦ Synopsis
The development of a highly sensitive method for the determination of nanomolar concentrations of hydrogen peroxide in the liquid phase is described. This paper demonstrates for the first time a flow injection analysis (FIA) system with immobilized enzyme reactor combined with a total internal reflective cell (a liquid waveguide capillary cell (LWCC)) and spectrophotometric detection, for the development of an improved procedure for the determination of hydrogen peroxide. Moreover, the newly synthesized 4-aminopyrazolone derivative, 4-amino-5-(p-aminophenyl)-1-methyl-2-phenyl-pyrazol-3-one (DAP), is used as a color coupler in its oxidative condensation with the sodium salt of N-ethyl-N-sulphopropylaniline sodium salt (ALPS) which acts as a hydrogen donor. Immobilization of peroxidase is achieved by coupling the periodate-treated enzyme to aminopropyl controlled-pore glass (CPG) beads. The determination of hydrogen peroxide is carried out in a 0.1 M phosphate buffer and the product is monitored at 590 nm with a charge-coupled device (CCD) detector equipped with fiber optics in a fully computerized system. The interference of different species, mainly ionic, was investigated.
The method permits detection down to 4 nmol l -1 hydrogen peroxide (signal-to-noise ratio = 3). A linear calibration graph was obtained over the range 20-700 nmol l -1 . The relative standard deviation (R.S.D.) at 300 nmol l -1 H 2 O 2 is 1.7% (n = 7). The method was successfully applied for the determination of hydrogen peroxide in samples from a vat-cleaning process.
📜 SIMILAR VOLUMES
A spectrophotometric, continuous flow-injection, merging zones manifold including an enzyme reactor was used for the determination of magnesium ion in serum by exploiting the activating effect of MgfII) on the hydrolysis of o-nitrophenyl-B-o-galactopyranoside by j?-o-galactosidase immobiliied on con
A flow-mJectlon system IS proposed for the determrnatlon of meat freshness, based on the simultaneous deternunation of polyammes and hypoxanthme A putrescme oxldase reactor and xanthme oxldase reactor were Incorporated at fured positrons m a flow system, which was based on the sphttmg of the flow af