A flow-mqectlon system for the chemdummometnc determmatlon of o-glucose m serum with an lmmobdlzed pyranose oxldase reactor IS described Pyranose oxldase IS lmmobdlzed on controlled-pore glass beads and packed mto a stamless-steel column Serum (100 nl) IS dduted lOWfold with water Sample solution (1
Spectrophotometric determination of magnesium in serum by using a flow-injection system with an immobilized enzyme reactor
✍ Scribed by J.M. Fernández-Romero; M.D.Luque de Castro; M. Valcárcel; R. Quiles-Zafra
- Publisher
- Elsevier Science
- Year
- 1993
- Tongue
- English
- Weight
- 446 KB
- Volume
- 283
- Category
- Article
- ISSN
- 0003-2670
No coin nor oath required. For personal study only.
✦ Synopsis
A spectrophotometric, continuous flow-injection, merging zones manifold including an enzyme reactor was used for the determination of magnesium ion in serum by exploiting the activating effect of MgfII) on the hydrolysis of o-nitrophenyl-B-o-galactopyranoside by j?-o-galactosidase immobiliied on controlled-pore glass. The advantageous features of the proposed method (determination range 5-20 pmol l-l, within-run and between-run relative standard deviations of 0.78-2.91% and 1.7-3.21%, respectively) allow the determination of the analyte in serum samples with good recoveries (%-113%). The results obtained were consistent with those provided by the conventional atomic absorption spectrometric method.