A flow injection system for assays of total cholesterol and free cholesterol was described. The total cholesterol assay system included an amperometric peroxidase electrode to measure hexacyanoferrate(III) converted from hydrogen peroxide, which was generated by injecting a 2-microliter sample into
Amperometric flow-injection method for determination of biogenic diamines and hypoxanthine by combined use of immobilized enzyme reactors and a peroxidase electrode
โ Scribed by Toshio Yao; Masahiro Satomura; Tamotsu Wasa
- Publisher
- Elsevier Science
- Year
- 1992
- Tongue
- English
- Weight
- 430 KB
- Volume
- 261
- Category
- Article
- ISSN
- 0003-2670
No coin nor oath required. For personal study only.
โฆ Synopsis
A flow-mJectlon system IS proposed for the determrnatlon of meat freshness, based on the simultaneous deternunation of polyammes and hypoxanthme A putrescme oxldase reactor and xanthme oxldase reactor were Incorporated at fured positrons m a flow system, which was based on the sphttmg of the flow after sample mlectlon and subsequent confluence before reaching the peroxldase electrode Because each channel has a different residence time, two peaks were obtained The first peak corresponded to the total polyamme concentration (putrescme, cadaverme and spermlchne) the second peak to hypoxanthme The cahbratlon graphs were linear m the range 1 x 10W6-5 x 10m4 M The detection limit was 0 5 x 10m6 M for polyammes and hypoxantlune The measurement of polyammes and hypoxanthme contents m porcine lom meat could be performed at a rate of 25 samples per hour with satlsfactoIy preclslon (< 1 5% R S D )