โฆ Synopsis
A robotic workstation for automated organic synthesis was developed. The system consists of a computer, robot, and several devices (reaction device, separation device, dispense device, etc.). The reaction device has four temperature-regulation units, each of which can freely change its set reaction temperature. Since each of the four can handle four reaction containers, 16 reactions under different conditions can be done simultaneously. The range of reaction temperature is from โซ03ืโฌะC to 160ะC. Reactions beyond the boiling points of solvents are possible by using reflux condensers. The separation processing device automatically performs solution extraction operations based on the detected liquid levels and interface positions. The dispensing device automatically pumps various kinds and amounts of reagents and/or solvents at various speeds into the reaction containers by switching valves through the digital syringe pump. This system can automate all synthesis and analysis process and can perform at least 3000 experiments (3000 containers) per year. Since its completion in 1996, it has conducted over 1500 experiments of various kinds.
๐ SIMILAR VOLUMES
This article describes a system capable of performing and analysing multiple degradation experiments on drug substances as part of the process of developing stability indicating separations methods. Qualitative data are generated on the signiยฎcant primary degradation processes of the drug of interes
We have further developed our microscale automated chemistry workstation for investigating the optimization of reaction conditions. The workstation includes a Cartesian robot for solvent delivery and reagent/sample handling, a 60-vessel reaction block, a UV-Vis absorption spectrometer, and vial rack
A general optimization methodology for the optimal design of robotic manipulators is presented and illustrated by its application to a realistic and practical three-link revolute-joint planar manipulator. The end-effector carries out a prescribed vertical motion for which, respectively, the average