Cramér–Rao bounds for three-point decomposition of water and fat
✍ Scribed by Angel R. Pineda; Scott B. Reeder; Zhifei Wen; Norbert J. Pelc
- Publisher
- John Wiley and Sons
- Year
- 2005
- Tongue
- English
- Weight
- 901 KB
- Volume
- 54
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
The noise analysis for three‐point decomposition of water and fat was extended to account for the uncertainty in the field map. This generalization leads to a nonlinear estimation problem. The Crámer–Rao bound (CRB) was used to study the variance of the estimates of the magnitude, phase, and field map by computing the maximum effective number of signals averaged (NSA) for any choice of echo time shifts. The analysis shows that the noise properties of the reconstructed magnitude, phase, and field map depend not only on the choice of echo time shifts but also on the amount of fat and water in each voxel and their alignment at the echo. The choice of echo time shifts for spin‐echo, spoiled gradient echo, and steady‐state free precession imaging techniques were optimized using the CRB. The noise analysis for the magnitude explains rough interfaces seen clinically in the boundary of fat and water with source images obtained symmetrically about the spin‐echo. It also provides a solution by choosing appropriate echo time shifts (−π/6 + π__k,__ π/2 + π__k,__ 7π/6 + π__k__), with k an integer. With this choice of echo time shifts it is possible to achieve the maximum NSA uniformly across all fat:water ratios. The optimization is also carried out for the estimation of phase and field map. These theoretical results were verified using Monte Carlo simulations with a newly developed nonlinear least‐squares reconstruction algorithm that achieves the CRB. Magn Reson Med, 2005. © 2005 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
We have derived analytical expressions of the Cramer-Rao lower bounds on spectral parameters for singlet, doublet, and triplet peaks in noise. We considered exponential damping (Lorentzian lineshape) and white Gaussian noise. The expressions, valid if a sufficiently large number of samples is used,
## Abstract An enhancement to Dixon's technique is described which can provide error‐free decomposition of water and fat proton images even in the presence of off‐resonance conditions which result from susceptibility differences, demagnetization, or shim errors. The method uses three measurements w
## Abstract To separate water and lipid resonance signals by phase‐sensitive MRI, a two‐point Dixon (2PD) reconstruction is presented in which phase‐unwrapping is used to obtain an inhomogeneity map based on only in‐phase and out‐of‐phase image data. Two relaxation‐weighted images, a “water image”
## Abstract Robust fat suppression techniques are required for many clinical applications. Multi‐echo water‐fat separation methods are relatively insensitive to B~0~ field inhomogeneity compared to the fat saturation method. Estimation of this field inhomogeneity, or field map, is an essential and