An investigation was performed to study the mechanical performance of fiber-reinforced composite hip prostheses in a femur. The main objective of the study was to evaluate the effect of fiber orientation of a composite femoral implant on the response of the surrounding femoral bone. A three-dimensio
Composite hip prosthesis design. II. Simulation
โ Scribed by Yildiz, Hasan ;Chang, Fu-Kuo ;Goodman, Stuart
- Publisher
- John Wiley and Sons
- Year
- 1998
- Tongue
- English
- Weight
- 624 KB
- Volume
- 39
- Category
- Article
- ISSN
- 0021-9304
No coin nor oath required. For personal study only.
โฆ Synopsis
An investigation was performed to study the mechanical performance of fiber-reinforced composite hip prostheses. In Part I of the study, a three-dimensional finite element code was developed for analyzing a composite hip prosthesis in a femur. The material properties of the composite were treated as anisotropic and inhomogeneous while the properties of the femoral bone were treated as anisotropic and homogeneous. All the materials were assumed to behave linear-elastically. Thermoplastic graphite/PEEK material was selected for the study. No slippage was assumed at the interface between the implant and the surrounding femoral bone. In Part II, numerical simulations were performed using the code to study the performance of a composite prosthesis in the femur. The stress/strain distributions, micromotions, and strain energy density of the surrounding femoral bone were evaluated and found to be related to initial fixation and long-term stability of the prosthesis in the femur. Numerous fiber orientations were studied, and the results of the calculations were compared with those generated from a prosthesis made of cobalt chrome and Ti-6Al-4V titanium alloys. Based on the analysis, it was shown that compared to conventional metallic implants more favorable stresses and deformations could be generated in the femur using composite implants. In addition, by changing fiber orientations according to femoral loads, a composite implant could be designed specifically for the left or the right femur.
๐ SIMILAR VOLUMES
## Abstract This investigation attempts to apply the concept of direct sketal attachment via tissue ingrowth to the attachment of femoral head prosthess to the femur. Stainless steel (316L) canine femoral head prostheses with a ceramic coated stem were implanted into the femora of dogs and their re
An accelerated cyclic loading corrosion test was used to determine the corrosion behavior of a commercial (GSP) and a prototype titanium hip prosthesis each with a modular neck. Four GSP and four prototype stems were subjected to a 2-Hz cyclic load ranging between 200 and 2,100 N for 1,000,000 cycle
## Abstract ## Background With the development of preoperative adjuvant treatment, imaging techniques, and improvement of surgical technique, limb salvage is now possible even in patients with pelvic tumors. However, reconstruction after periacetabular resection is complicated and challenging. ##