๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

Design-related fretting wear in modular neck hip prosthesis

โœ Scribed by Viceconti, M. ;Ruggeri, O. ;Toni, A. ;Giunti, A.


Publisher
John Wiley and Sons
Year
1996
Tongue
English
Weight
745 KB
Volume
30
Category
Article
ISSN
0021-9304

No coin nor oath required. For personal study only.

โœฆ Synopsis


An accelerated cyclic loading corrosion test was used to determine the corrosion behavior of a commercial (GSP) and a prototype titanium hip prosthesis each with a modular neck. Four GSP and four prototype stems were subjected to a 2-Hz cyclic load ranging between 200 and 2,100 N for 1,000,000 cycles. Three stems were tested in an environment of FeCl3 solution, three stems were tested in Ringer's solution, and two stems were tested in air. After cyclic loading, the specimens were carefully examined with optical and scanning electron microscopy (SEM). None of them showed macroscopic or microscopic signs of corrosion, regardless of the environment to which the specimens were subjected. However, macroscopic evidence of mechanical fretting was present at the neck-stem modular junction, primarily concentrated at the medial contact point between stem and neck, especially for the prototype stems. SEM analysis confirmed these observations. The appreciable differences observed between the two designs suggest that the problem can be minimized or eliminated with an accurately designed taper fitting.


๐Ÿ“œ SIMILAR VOLUMES


Fretting wear in a modular neck hip pros
โœ Viceconti, Marco ;Baleani, Massimiliano ;Squarzoni, Stefano ;Tonil, Aldo ๐Ÿ“‚ Article ๐Ÿ“… 1997 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 456 KB

In vitro cyclic load fretting tests were conducted on a prototype of a cementless, modular neck, hip prosthesis. The study had three major objectives: to determine the amount of fretted material in the tapered-neck joint under various load cycle amplitudes, to determine the fretting damage evolution