An accelerated cyclic loading corrosion test was used to determine the corrosion behavior of a commercial (GSP) and a prototype titanium hip prosthesis each with a modular neck. Four GSP and four prototype stems were subjected to a 2-Hz cyclic load ranging between 200 and 2,100 N for 1,000,000 cycle
Fretting wear in a modular neck hip prosthesis
β Scribed by Viceconti, Marco ;Baleani, Massimiliano ;Squarzoni, Stefano ;Tonil, Aldo
- Publisher
- John Wiley and Sons
- Year
- 1997
- Tongue
- English
- Weight
- 456 KB
- Volume
- 35
- Category
- Article
- ISSN
- 0021-9304
No coin nor oath required. For personal study only.
β¦ Synopsis
In vitro cyclic load fretting tests were conducted on a prototype of a cementless, modular neck, hip prosthesis. The study had three major objectives: to determine the amount of fretted material in the tapered-neck joint under various load cycle amplitudes, to determine the fretting damage evolution, and to determine the effect of different-sized stem bodies on the production of debris. All the tests produced some fretting microdamage on the tapered surface although the extent was quite different among test groups. The amount of abraded material increased almost linearly with the applied load magnitude but not with the number of load cycles. The amount of weight loss was higher in the large stem bodies than in the small ones. Weight loss ranged from 0.28 Β± 0.10 mg for small stem bodies loaded 5.5 million times up to 2300N to 2.54 Β± 0.53 mg for large stem bodies located 20 million times up to 3300N. Considering the large-size stem results, and assuming one million load cycles between 300N and 3300N to be the average yearly load history, the modular neck tapered joint would produce 0.6 mg/year of metal debris. The clinical impact of this observation is unknown; however, some of the literature on the presence of metal in patient tissues and fluids supports the hypothesis that a normal and stable prosthesis is likely to produce less than 10 mg/year of metal debris. Thus, a further production of 0.6 mg/year due to the modular neck should not have any significant effect.
π SIMILAR VOLUMES
## Abstract The use of multipleβcomponent systems in orthopedic surgery gives the surgeon increased flexibility in choosing the optimal implant, but introduces the possibility of interfacial corrosion. Such corrosion could limit the longevity of prostheses due either to tissue reactions to corrosio
Modular hip prostheses are increasing in variety and utilization. Component stability, high endurance limit, and minimal particulate debris generation are critical for long-term clinical success. The purpose of this study was to characterize the fatigue response and evaluate the in v i m potential f
## Abstract Wear of the polyethylene in total joint prostheses has been a source of morbidity and early device failure, which has been extensively reported in the last 20 years. Although research continues to attempt to reduce the wear of polyethylene jointβbearing surfaces by modifications in poly