Comparison of the performance of round and rectangular wire in small solenoids for high-field NMR
✍ Scribed by Yu Li; Andrew G. Webb; Saikat Saha; William W. Brey; Cherian Zachariah; Arthur S. Edison
- Publisher
- John Wiley and Sons
- Year
- 2006
- Tongue
- English
- Weight
- 235 KB
- Volume
- 44
- Category
- Article
- ISSN
- 0749-1581
- DOI
- 10.1002/mrc.1777
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
This paper considers the effects of conductor geometry on the performance of small solenoidal coils for high‐field NMR. First, a simple analytical model is presented for investigating the effects of conductor geometry on the current distribution in such coils. The model was used to derive optimum parameters for coils constructed from wire with either rectangular or circular cross‐sections as a function of the length‐to‐diameter ratio. Second, a commercial software package utilizing full three‐dimensional finite‐element solutions to Maxwell's equations was used to confirm the basic findings of the simple analytical model, and also to compare simulated S/N estimations with experimental NMR spectra acquired with 2.5 mm and 1.0 mm‐diameter solenoid coils: reasonable agreement was found. Third, as a demonstration of the usefulness of such coils for mass‐limited samples, multidimensional experiments were performed at 750 MHz on ∼4.7 nmol (41 µg) of PF1061, a protein from Pyrococcus furiosus. Copyright © 2006 John Wiley & Sons, Ltd.
📜 SIMILAR VOLUMES
High-performance capillary electrophoresis (HPCE) and high-performance liquid chromatography (HPLC) were developed and applied to the determination of acyclovir (ACV) and guanine (C) in pharmaceuticals. The comparison study showed that two methods gave comparable results in linear range, recovery an
A comparison has been made of high-performance liquid chromatographic and atomic absorption spectrometric methods for the determination of Fe(II1) and Al(II1) in soil and clay samples, following five different digestion/extraction schemes. Good correlations were obtained in the case of Fe(II1) deter
## High-performance liquid chromatography is used to develop a sensitive method for the determination of tocopherol levels in the plasma of cattle and pigs. This method is compared with a similar method using UV detection and one using direct fluorescence determination of tocopherol. Finally a do