A 82Sr/82Rb generator system is described which is shown to be suitable for continuous intravenous infusion in man. The breakthrough of the 82Sr parent has been closely monitored and remained less than 18.5 Bq mL-1 of infusate. A method using a 0.05% solution of sodium hypochlorite to disinfect the
Clinical uses of 82Sr/82Rb generators
โ Scribed by T. Jones
- Publisher
- Elsevier Science
- Year
- 1987
- Weight
- 327 KB
- Volume
- 38
- Category
- Article
- ISSN
- 0883-2889
No coin nor oath required. For personal study only.
๐ SIMILAR VOLUMES
Static (dose calibrator) assays of 82Rb bolus yield erroneous estimates of administered activity by ignoring the asymmetric output of 82Sr/82Rb generators. Though on-line monitoring of elution profile improves accuracy for quantitative purposes, it requires an elaborately shielded detector and assoc
The activity of 82Rb produced from a 82Sr/82Rb generator is dependent on elution conditions (volume and eluent flow rate) and sampling conditions (time and position of collection). Assays for 82Rb in generator eluates are described using a commercial dose calibrator in a static procedure and a plast