𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Model for 82Sr/82Rb generator elution profiles: A secondary approach to radioassay/dosimetry

✍ Scribed by Vijay Dhawan


Publisher
Elsevier Science
Year
1987
Weight
567 KB
Volume
38
Category
Article
ISSN
0883-2889

No coin nor oath required. For personal study only.

✦ Synopsis


Static (dose calibrator) assays of 82Rb bolus yield erroneous estimates of administered activity by ignoring the asymmetric output of 82Sr/82Rb generators. Though on-line monitoring of elution profile improves accuracy for quantitative purposes, it requires an elaborately shielded detector and associated electronics at patient-study site. Our alternative approach based on mathematical description of the dynamic characteristics of the generator accurately predicts elution profiles over a broad range of flow rates.