Abstract

Purpose

To check the bioeffects of the components of magnetic resonance imaging (MRI). MRI is based on an assumedly harmless interaction between certain nuclei in the body when placed in a strong magnetic field and radio wave fields. There are three key factors actuating on the examining body: a powerful static magnetic field (SMF), magnetic gradient fields (MGFs), and pulsed radiofrequency (RF) radiation.

Materials and Methods

In vitro cells (L‐132 cells) were used as biosensors, and different cellular compounds were used as biomarkers (heat shock proteins [HSPs] and their messenger ribonucleic acids [mRNAs], calcium, and adenosine‐3′,5′‐cyclic monophosphate [cAMP]). The biosensors were placed in the bore of a 1.5‐T MRI machine and the different electromagnetic fields were operated.

Results

HSPs and their mRNAs and cAMP did not respond to SMF, MGFs, or RF radiation. RF radiation increased cytosolic calcium concentration (18%, P < 0.05).

Conclusion

Although MRI procedures do not induce any cellular stress response, it may cause an unfathomable calcium increase in vitro. Although the in vitro experimental conditions are not totally comparable to clinical situations, the usefulness of the in vivo biological dosimetry, circulating leukocytes as biosensors, and HSPs and/or calcium as biomarkers is suggested. J. Magn. Reson. Imaging 2002;15:584–590. © 2002 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES


Magnetic resonance imaging of perfusion
✍ D. Le Bihan 📂 Article 📅 1990 🏛 John Wiley and Sons 🌐 English ⚖ 620 KB

## Abstract Recent developments have shown that diffusion and blood microcirculation (perfusion) could be imaged and measured noninvasively by MRI. The purpose of this presentation is to overview the different approaches that use __B__~0~ field gradients to monitor diffusion / perfusion. The princi

Transesophageal magnetic resonance imagi
✍ Kendrick A. Shunk; Joao A.C. Lima; Alan W. Heldman; Ergin Atalar 📂 Article 📅 1999 🏛 John Wiley and Sons 🌐 English ⚖ 238 KB 👁 2 views

The purpose of this study was to develop a non-invasive method of imaging the thoracic aorta that would provide both morphological detail within the aortic wall and information about regional aortic wall motion. An esophageal probe is described that allows transesophageal MR imaging (TEMRI) of the t