𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Transesophageal magnetic resonance imaging

✍ Scribed by Kendrick A. Shunk; Joao A.C. Lima; Alan W. Heldman; Ergin Atalar


Publisher
John Wiley and Sons
Year
1999
Tongue
English
Weight
238 KB
Volume
41
Category
Article
ISSN
0740-3194

No coin nor oath required. For personal study only.

✦ Synopsis


The purpose of this study was to develop a non-invasive method of imaging the thoracic aorta that would provide both morphological detail within the aortic wall and information about regional aortic wall motion. An esophageal probe is described that allows transesophageal MR imaging (TEMRI) of the thoracic aorta and has several potential advantages over the competing non-vasculoinvasive techniques of transesophageal echocardiography (TEE) or standard MRI. The probe consists of a loopless antenna housed inside a modified Levin gastric tube, with external matching and tuning circuitry. Using this probe, the thoracic aorta has been imaged in longitudinal and cross-sectional views. Details of the aortic wall were readily seen. Tissue tagging for measurement of focal stress/strain relationships was demonstrated to be feasible. TEMRI avoids the risks inherent in intravascular MRI yet provides comparable image quality. Potential applications of the device are discussed.


πŸ“œ SIMILAR VOLUMES


Understanding Magnetic Resonance Imaging
✍ E. Mark Haacke, πŸ“‚ Article πŸ“… 1999 πŸ› John Wiley and Sons 🌐 English βš– 22 KB πŸ‘ 2 views
Human magnetic resonance imaging at 8 T
✍ P.-M. L. Robitaille; A. M. Abduljalil; A. Kangarlu; X. Zhang; Y. Yu; R. Burgess; πŸ“‚ Article πŸ“… 1998 πŸ› John Wiley and Sons 🌐 English βš– 174 KB πŸ‘ 1 views

In this work, we present the first human magnetic resonance image (MRI) obtained at ultrahigh field strengths (8 T). We demonstrate that clinical imaging will be possible at 8 T and that reasonable quality head images can be obtained at this field strength. Most importantly, we emphasize that the po