## Abstract A method of correcting motional artifacts in diffusion‐weighted images is described. Motion causes changes in the phase of the __k__‐space NMR signal and thereby introduces positional shifts (or ghosts) in the spatial domain. By correcting the phase of the NMR signal before Fourier tran
Analysis and correction of motion artifacts in diffusion weighted imaging
✍ Scribed by Adam W. Anderson; John C. Gore
- Publisher
- John Wiley and Sons
- Year
- 1994
- Tongue
- English
- Weight
- 903 KB
- Volume
- 32
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
For diffusion‐weighted magnetic resonance imaging and under circumstances where patient movement can be modeled as rigid body motion, it is shown both theoretically and experimentally that translations and rotations produce phase errors which are zero‐ and first‐order, respectively, in position. Whlile a navigator echo can be used to correct the imaging data for arbitrary translations, only when the diffusion gradient is applied in the phase encode direction is there sufficient information to correct for rotations around all axes, and therefore for general rigid body motion. Experiments in test objects and human brain imaging confirm theoretical predictions and demonstrate that appropriate corrections dramatically improve image quality in vivo.
📜 SIMILAR VOLUMES
## Abstract Patient motion and image distortion induced by eddy currents cause artifacts in maps of diffusion parameters computed from diffusion‐weighted (DW) images. A novel and comprehensive approach to correct for spatial misalignment of DW imaging (DWI) volumes acquired with different strengths
## Abstract ## Purpose: To provide a quantitative assessment of motion and distortion correction of diffusion‐weighted images (DWIs) of the breast and to evaluate the effects of registration on the mean apparent diffusion coefficient (mADC). ## Materials and Methods: Eight datasets from four pat
## Abstract Nonuniformities of magnetic field gradients can cause serious artifacts in diffusion imaging. While it is well known that nonlinearities of the imaging gradients lead to image warping, those imperfections can also cause spatially dependent errors in the direction and magnitude of the di
## Abstract ## Purpose To evaluate the performance of the autofocusing (AF) motion correction technique in high‐resolution trabecular bone imaging where image signal‐to noise ratio (SNR) is limited. ## Materials and Methods Raw data from 26 clinical three‐dimensional (3D) wrist exams were motion