𝔖 Bobbio Scriptorium
✦   LIBER   ✦

A clinical evaluation of the Bonfils Intubation Fibrescope

✍ Scribed by M. Halligan; P. Charters


Publisher
John Wiley and Sons
Year
2003
Tongue
English
Weight
118 KB
Volume
58
Category
Article
ISSN
0003-2409

No coin nor oath required. For personal study only.

✦ Synopsis


Summary

The Bonfils Intubation Fibrescope is a rigid optical instrument for performing orotracheal intubation. We describe its introduction into our clinical practice in 60 patients with normal airways who required orotracheal intubation for elective surgery. Two anaesthetists each performed 30 attempts to intubate, in turn, in patients who received a standard general anaesthetic with neuromuscular blockade. Intubation was successful in 59 out of 60 cases. The median (IQR [range]) time to intubation was 33 s (24–50 [13–180] s). Median (IQR [range]) verbal rating score for difficulty was 2 (1–3 [0–10]). There was a significant correlation between the intubation times and the verbal rating score (p < 0.01). There was evidence of airway trauma in the single patient in whom intubation failed. The Bonfils Intubation Fibrescope is an effective instrument for orotracheal intubation in normal subjects.


πŸ“œ SIMILAR VOLUMES


Effect of cricoid pressure on the ease o
✍ T. Asai; K. Murao; S. Johmura; K. Shingu πŸ“‚ Article πŸ“… 2002 πŸ› John Wiley and Sons 🌐 English βš– 198 KB

**Summary** We studied the reasons for difficulty with tracheal intubation over a fibrescope, and whether cricoid pressure facilitated intubation, in 50 patients. After induction of anaesthesia and neuromuscular blockade in 10 patients (phase 1), we attempted to clarify the reasons for difficulty in

Use of the bougie in simulated difficult
✍ K. C. Hames; J. J. Pandit; A. G. Marfin; M. T. Popat; S. M. Yentis πŸ“‚ Article πŸ“… 2003 πŸ› John Wiley and Sons 🌐 English βš– 83 KB

## Summary We studied the success rates for tracheal intubation in 64 healthy patients during simulated grade III laryngoscopy after induction of anaesthesia, using either the single‐use bougie or oral flexible intubating fibrescope, both in conjunction with conventional Macintosh laryngoscopy. Pat

A comparison of blind and lightwand-guid
✍ S. Kihara; S. Watanabe; N. Taguchi; A. Suga; J. R. Brimacombe πŸ“‚ Article πŸ“… 2000 πŸ› John Wiley and Sons 🌐 English βš– 165 KB

We have tested the hypothesis that intubation success rates, haemodynamic changes, airway complications and postoperative pharyngolaryngeal morbidity differ between blind and lightwand-guided intubation through the intubating laryngeal mask airway. One hundred and twenty paralysed anasthetised adult

A clinical evaluation of the B-D Pen
✍ Rowe, B R ;Pizzey, M ;Barnett, A H πŸ“‚ Article πŸ“… 1992 πŸ› John Wiley and Sons 🌐 English βš– 214 KB

## Abstract The applicability and acceptability of the B‐D pen device (Becton Dickinson & Co) with Humulin insulin cartridges (Eli Lilly & Co., Ltd) in the daily treatment of one hundred diabetic patients were investigated during a three‐month open study. Questionnaire ratings demonstrated a high d

A comparison of the intubating laryngeal
✍ D. N. Lucas; S. M. Yentis πŸ“‚ Article πŸ“… 2000 πŸ› John Wiley and Sons 🌐 English βš– 120 KB

We conducted a prospective randomised study to compare the intubating laryngeal mask tracheal tube with a standard tube, for ease of tracheal intubation over a fibrescope. Thirty‐six patients were investigated, using a sequential analysis technique. Ease of intubation was improved when the intubatin

Clinical evaluation of asaley
✍ Bodey, Gerald P. ;Gottlieb, Jeffrey A. ;Burgess, M. Andrew ;Alexanian, Raymond πŸ“‚ Article πŸ“… 1977 πŸ› John Wiley and Sons 🌐 English βš– 435 KB

## Abstract Asaley is an L‐leucine derivative of sarcolysin which is more active against some rodent tumors. Studies in the USSR demonstrated activity in patients with ovarian and breast carcinoma, Hodgkin's disease, and multiple myeloma. This study in 73 evaluable patients indicated that an approp